Understanding Test Results and Messages
About Quality Control Testing
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the
Control Solutions GC700 when testing your
Blood Glucose
Monitoring System GM700S in the Control Solution Mode. If the test result is within the Control
Solution Range printed on the strip vial label, the
Blood Glucose Monitoring System
GM700S passes the Quality Control Test and your
Blood Glucose Monitoring System
GM700S is working properly.
Control Solution Range:
Example of Control Solution Range
printed on your test strip vial label.
The
Meter
displays results between
10~600 mg/dL (0.6~33.3 mmol/L)
If the test result is below
, “ ” will appear on the
screen. Please repeat your test with by a new test strip.
If you still get a “ ” result, contact your healthcare professional.
If the test result is above
“ ” will appear on the
screen. Please repeat your test again with a new test strip.
If you still get a “ ” result, contact your healthcare professional.
Blood Glucose
GM700S
10 mg/dL (0.6 mmol/L)
600 mg/dL (33.3 mmol/L),
Control Solution Range mg/dL
Low: 20-48
Normal: 107-145
High: 315-426
NOTE
- If your blood glucose result is unusually high or low, or if you question your testing results,
repeat the test with a new test strip. You can also run a Quality Control Test to check your
meter and test strip. If the test result still remains unusually high or low, contact your
healthcare professional immediately.
- If you are experiencing symptoms that are not consistent with your blood glucose test
results and you have made sure to follow all instructions of this manual, contact your
healthcare professional immediately.