About Quality Control Testing

About Quality Control Testing

CAUTION

- Each time you open a new bottle of control solution, write the expiration date on the label. 

The

Control Solution GC700 is good for 3 months after opening the bottle, or 

until the expiration date printed on the label, whichever comes first.

                 

- Wipe the bottle cap with a clean tissue before tightly closing the bottle of control solution.
- Close the bottle of

 tightly immediately after using.

- Check the expiration date before use (Refer to the

Control Solution GC700 

Package Insert).

- Keep control solution bottles out of reach of children.

                 

                 

Control Solution GC700

Example

Exp. Date 2017-10

When should a Quality Control Test Be Performed ?

- To ensure that your

Blood Glucose Meter

 and

Blood Glucose Test 

Strip 

 are working properly.

- To confirm that you are following the correct testing procedures.
- To prepare for your initial blood glucose test.
- To check the

Blood Glucose Test Strip 

 when you open a new vial of strips.  

- To practice testing.

Required Items for Quality Control Tests
To perform a quality control test, prepare the items below :

Blood Glucose Meter

 

Blood Glucose Test Strips 

-

  

Control Solution GC700

                 

                 

                 

                 

                
                

                

 GM700S

GS700

GS700

 GM700S

 GM700S

GS700

To check your

Blood Glucose Meter

 after it has been dropped, damaged or 

exposed to liquids.

- If you suspect that your test results are inaccurate, or if your test results are not consistent with the 

way you feel.

 
 

Summary of Contents for GM700S

Page 1: ...outh Dist Taichung City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E mail info bionime com BIONIME GmbH Tramstrasse 16 9442 BerneckSwitzerland E mail info bionime ch Blo...

Page 2: ......

Page 3: ...Blood Glucose Monitoring System GM700S is manufactured and supported by Bionime Corporation If you have any questions or concerns please contact your local PharmEvo Customer Service or email to evoche...

Page 4: ...trol test regularly to make sure the test results are accurate GM700S GS700 Other brands test strips should not be used under any circumstances The use of other brands strips may give inaccurate resul...

Page 5: ...nformation Meter About the Blood Glucose Monitoring System The Blood Glucose Monitoring System The Blood Glucose Meter The Blood Glucose Test Strip Before Testing Meter Activation and Battery Change S...

Page 6: ...rip GS700 Package Insert 4 Control Solution GC700 Insert 5 Blood Glucose Meter GM700S with 1 CR2032 battery installed 6 Blood Glucose Test strips GS700 10 25 pcs 7 Control Solution GC700 Normal Level...

Page 7: ...GM700S 11 10 The Blood Glucose Meter GM700S Right Button Battery Cover Meter Battery The Blood Glucose Meter GM700S Left Button Main Button Display Window Test Strip Port 2D barcode with serial number...

Page 8: ...esult Warns when the batteries are low or must be replaced Indicates when to apply the blood sample Appears after you insert test strip into meter Appears when an error occurs Current date under time...

Page 9: ...4 30 C or 39 86 F and dry location 90 relative humidity Do not expose to direct sunlight or heat If the Blood Glucose Meter GM700S and Test Strips are exposed to a temperature outrange environment ple...

Page 10: ...lease follow the local regulation to properly discard a used battery 4 The Blood Glucose Meter 5 Press any button to exit the self test and enter Setting Mode GM700S replacing the battery all symbols...

Page 11: ...this time the screen will go blank until you hear a beep After the beep the meter will turn on into the Setting Mode The display screen will show setting data GM700S GM700S Setting Up Your Meter Setti...

Page 12: ...ttings during Meter Set up for over 2 minutes the Blood Glucose Meter GM700S will leave setting mode and power off automatically Your Meter default unit is set as mg dL Setting Up Your Meter Setting t...

Page 13: ...utes if no buttons are pressed or no strip is inserted Turning On Off the Meter Handling the Blood Glucose Test Strip GS700 How to handle the ood Glucose Test Strip Inserting the Blood Glucose Test St...

Page 14: ...f the test strip port Please follow your local regulations and discard used strips properly GS700 1 2 3 Getting Ready for Testing Before performing a blood glucose test prepare the items below Blood G...

Page 15: ...r thick or calloused skin 1 2 3 4 5 6 Performing a Blood Glucose Test 7 Hold the base in one hand and pull on the plunger with the other hand The device will be cocked Release the plunger and it will...

Page 16: ...p and repeat with a new Blood Glucose Test Strip GS700 15 Countdown Mode will begin on the display window After 5 seconds your test result will appear 13 15 Performing a Blood Glucose Test 14b 14c 14a...

Page 17: ...ious diseases even after you have performed the cleaning and disinfecting procedure Users should wash hands thoroughly with soap and water after handling the meter lancing device and test strips Pleas...

Page 18: ...a test strip Do not use expired Test Strips Use each Test Strip immediately after removing it from the vial Do not reuse Test Strips Apply the blood sample only on the sample port of the test strip D...

Page 19: ...pear on the screen Please repeat your test with by a new test strip If you still get a result contact your healthcare professional If the test result is above will appear on the screen Please repeat y...

Page 20: ...trol Solution GC700 Example Exp Date 2017 10 When should a Quality Control Test Be Performed To ensure that your Blood Glucose Meter and Blood Glucose Test Strip are working properly To confirm that y...

Page 21: ...ontrol solution This could contaminate the meter via the test strip port Keep the test strip port clean and dry Clean immediately if the test strip port is stained or is overly exposed to moisture Do...

Page 22: ...C700 The control solution test result will not be used when recalling test average results Your control solution test results should fall within the control solution range If the results are within th...

Page 23: ...i results control solution test results and test results under abnormal temperature 43 F 111 F will be excluded from average calculations The Average function requires that the correct time and date i...

Page 24: ...r test strip port is stained with blood control solution or any liquid please use a dry tissue or alcohol swab to clean it immediately Do not use anything wet to clean it Perform a quality control tes...

Page 25: ...Changing the Battery The meter should work normally after finishing the above steps If the meter still does not work please contact your local customer service GM700S Pharmevo Strip Error Sampling Err...

Page 26: ...smolar patients Regarding potential electromagnetic or other interference keep meter out of electromagnetic radiation sources CAUTION A blood sample should ONLY be applied to the test strip after the...

Page 27: ...tions Measurement Technology Sample Minimum Sample Volume Measuring Range Test Time Memory Capacity Power Saving Operating Temperature Operating Relative Humidity Dehydrogenase Electrochemical Sensor...

Page 28: ...e from defects in materials and workmanship for five years from the date of purchase GM700S GM700S Different models have different specifications This warranty applies only to the Blood Glucose Monito...

Page 29: ...r SteriLance Medical SuZhou Inc No 68 LiTangHe RD XiangCheng SuZhou JiangSu 215133 P R China Product complies with Medical Device Directive 93 42 EEC CE 0197 EC Rep EMERGO EUROPE Molenstraat 15 2513 B...

Reviews: