View Window Appearance
View Window Appearance
Make sure your blood sample covers the
whole area of the view window to get an
accurate test result. An insufficient blood
sample will result in an error message
("Er4"). If this occurs, repeat the test with a
new test strip.
Insufficient blood sample
Sufficient blood sample
CAUTION
- Check the expiration date printed on the strip vial every time you use a test strip. Do not use
expired Test Strips.
- Use each Test Strip immediately after removing it from the vial.
- Do not reuse Test Strips.
-
- Apply the blood sample only on the sample port of the test strip.
Do not perform testing immediately after moving from one ambient temperature to another
(e.g. after coming indoors from the outside). Allow 30 minutes for the temperature of the
meter and the test strips to equilibrate.
- Please don't drip or inject the blood sample directly by syringe to the sample
entry of test strip. Doing this might contaminate the meter or cause damages
and is not recommended.