PF9800 series User
’
s Manual
Copyright
©
EVERFINE, Copy or spread without authorization is prohibited.
9
12 months after calibration
Temperature : 23
℃±
5
℃
Humidity: 30% to 75% R.H.
Supply voltage: 220V
±
10V
Input waveform: 45Hz~65Hz sine
wave
Input range: (10%~100%)Range
Common
mode voltage:
0VDC
Power and PF are measured under
the voltage of 110/220/380V
±
(0.4% of rdg + 0.1% of
rng +1 digit)
±
(0.004+0.001
of rdg + 1
digit)
±
(0.1%of rdg+1 digit)
(3) Harmonic measurement(PF9810/PF9811)
Table 2.3
Function of harmonic measurement
Method
Phase-locked loop synchronization method(PLL)
Analysis of
Projects
The effective value of each harmonic component and relative value of
voltage, current and active power
The relative values of total harmonic distortion of the voltage and current
The calculation method of harmonic distortion is IEC
the total calculation method of harmonic distortion can be set IEC or CSA in
computer.
Frequency range
The fundamental frequency of PLL source is between45Hz and 65Hz.
PLL Source
The input signal o should be more than 50% of the range.
Accuracy
±
(0.3% of rng +5% of rdg)
Analysis Order
1~50
Update Interval of
harmonic display
Approx.2 seconds once time
(4) Communication(PF9810/PF9811)
Table 2.4
Communication settings
Communication interface
Standard Configuration:RS-232,Optional:RS-485
、
LAN
Baud rate
2400
Communication address
0, 1, 2
……
31
2.2 General technical requirements
Table 2.5
General technical requirements
Preheating time
≥
30 minutes
Operating
Temperature: 5
℃~
40
℃