33
11
SPECIFICATIONS
TDS Range
Resolution
Cell Constant
0.00 – 10.00 ppm
0.01 ppm
0.1
0.0 – 100.0 ppm
0.1 ppm
0.1
0.0 – 100.0 ppm
0.1 ppm
1
0 – 1000 ppm
1 ppm
1
0.00 – 5.00 ppt
0.01 ppt
1
0.00 – 10.00 ppt
0.01 ppt
1
0.0 – 100.0 ppt
0.1 ppt
1
0.0 – 200.0 ppt
0.1 ppt
10
Relative Accuracy
± 1 % of full scale
Temperature
-10 to 110
o
C
Resolution/Accuracy
0.1 / ± 0.5
o
C
Sensor
Pt 100 (3-wire)
Temperature Comensation
Automatic / Manual (0 to 50
o
C)
Temperature Coefficient
0.0 to 5.0 %
TDS Factor
0.40 to 1.00
Set-point And Controller Functions
Function Limit
Control
Switching Conductivity Hysteresis
5% of full scale
Contact Outputs, Controller
2 SPDT relays
Switching Voltage / Current / Power
Max 250 VAC / Max 3A / Max 600 VA
Electrical Data And Connections
Power Requirements
85 to 260 VAC, 50/60 Hz or DC
Signal Output / Load
4-20 mA galvanically isolated / 200
Ω
Connection Terminals
4 Detachable connectors
(3-pin; 7-pin & 9-pin terminal blocks)
Main Fuse
250 mA, Anti-surge (BUSSMAN S504+250mA)
Environmental Conditions
Ambient Temp. Operating Range
- 10 to 50
o
C (14 to 122
o
F)
Rel. Humidity
10 to 95 % (non-condensing)
Mechanical Specifications
Dimensions (Panel Housing – W x H x D)
96 x 48 x 150 mm
Weight
300g (350g boxed)