24
8.4
Set Point 2 – P2.0
Setup program P2.0 allows you to set parameters for relay 2.
P2.1: select relay 2 set point value
P2.2: select relay 2 as low or high set point
P2.3: set relay 2 hysteresis value (dead band)
P
2.0
P
2.1
P
2.2
P
2.3
1800
100
HYS
t
SP 2
SP 2
SP 2
LOW
HIGH
Press
▲
and
▼
keys together (ESCAPE) at anytime, to leave Setup mode and return to
Measurement mode.
P2.1: Select relay set point value
Set the TDS value that will activate Relay 2. If measured value overshoots or
undershoots Set Point value, Relay 2 will be activated and corresponding LED on front
panel will light.
1.
Key in password “022” as per procedure in Section 5.
2.
Press
▲
key until screen displays P2.0 and SP2. Press ENTER.
3.
The screen will scroll P2.1, SP2, then current set point value.
4.
Press
▲
or
▼
keys and adjust second relay set point. Default value is 90% of full
scale of range selected.
5.
Press ENTER to confirm and continue to step three of P2.2, or press
▲
and
▼
keys together to return to P2.0. Press
▲
and
▼
keys together to return to TDS
measurement mode.