- 25 -
9.7
Viewing pH Electrode Data – P5.0
In pH mode, Program 5 has three “view only” options that lets you check the
electrode parameters for diagnostic purposes.
P5.1: view pH electrode offset
P5.2: view pH electrode slope
P5.3: view temperature probe offset (ATC on only)
These parameters will change each time you recalibrate the controller.
P5.0
elec
Slpe
P5.1
P5.2
P5.3
p.ofs
t.ofs
0.0
5 .
92
5
.
0
Press
▲
and
▼
keys together (ESCAPE) at anytime, to leave the Setup mode
and return to Measurement mode.
P5.0: Viewing pH electrode data
1.
Key in the password “022” as per procedure in Section 5.
2.
Press
▲
key four times. The screen will scroll P5.0 and ELEC. Press
ENTER.
3.
The screen will scroll P5.1, then P.OFS, then pH electrode offset in mV at
the pH 6.86 / 7.00 calibration point.
4.
Press ENTER. The screen will display P5.2, then SLPE, then the electrode
slope in mV.
5.
MTC on: press ENTER to return to general Setup mode.
ATC on: go to next step.
6.
The screen will scroll P5.3, then T.OFS, then the temperature offset value.
7.
Press ENTER to return to general Setup mode.
Press
▲
and
▼
keys together (ESCAPE) to return to measurement mode.