MM850592 ed. 01 - Mini MMX II
5
SPECIFICATIONS
The instrument complies with
: EN61010-1.
Insulation
: Class2, Double insulation.
Overvoltage category
: CATIII 1000V, CATIV600V.
Display
: 4000 counts LCD display with function indication.
Polarity
: Automatic, (-) negative polarity indication.
Overrange
: “OL” mark indication.
Low battery indication
: The “BAT” is displayed when the battery voltage drops below the operating level.
Measurement rate
: 2 times per second, nominal.
Auto power off
: Meter automatically shuts down after approx. 15 minutes of inactivity.
Operating environment
: 0°CC to 50°C (32°F to 122°F) at < 70 % relative humidity.
Storage temperature
: -20°C to 60°C (-4°F to 140°F) at < 80 % relative humidity.
For inside use, max height
: 2000m
Pollution degree
: 2
Power
: One 9V battery , NEDA 1604, IEC 6F22.
Dimensions
: 182 (H) x 82 (W) x55 (D) mm
Weight
: Approx.: 375g.
Accuracy is given at 18°C to 28°C (65°F to 83°F), less than 70 % RH
DC Voltage
(Auto-ranging)
Range Resolution
Accuracy
400.0mV 0.1mV
+0.5% of rdg + 2 dgts
4.000V 1mV
40.00V 10mV
400.0V 100mV
+1.2% of rdg + 2 dgts
1000V 1V
+1.5% of rdg + 2 dgts
Input Impedance: 7.8M
Ω
.
Maximum Input: 1000V dc or 1000V ac rms.
AC Voltage
(Auto-ranging except 400mV)
Range Resolution
Accuracy
400.0mV 0.1mV +1.5%of rdg + 70 dgts
4.000V 1mV +1.2% of rdg + 3 dgts
40.00V 10mV
400.0V 100mV
+1.5% of rdg + 3 dgts
1000V 1V +2.0% of rdg + 4 dgts
Input Impedance: 7.8M
Ω
.
AC Response: 50 Hz to 400 Hz
Maximum Input: 1000V dc or 1000V ac rms.
DC Current
(Auto-ranging for
µ
A and mA)
Range Resolution
Accuracy
400.0
µ
A 0.1
µ
A
+1.0% of rdg + 3 dgts
4000
µ
A 1
µ
A
40.00mA
10
µ
A
400.0mA
100
µ
A
+1.5% of rdg + 3 dgts
10A 10mA +2.5% of rdg + 5 dgts
Overload Protection: 0.5A / 1000V and 10A / 1000V Fuse.
Maximum Input: 400mA dc or 400mA ac rms on
µ
A / mA ranges, 10A dc or ac rms on 10A range.