2408
i Indicator Engineering Manual
3.3.5.2
Calibration Type = Load Cell
(
<
tYPE
>
=
<
Ld.C
>
)
See also section 3.6.3.
Mnem-
onic
Meaning
Adjustable Range
Default
setting
Customer
setting
tArE
Performs automatic ‘Tare’ correction
See ’User Calibration’ section for further
description
OFF
= Off
OFF
on
= Start correction
busy
= Calculating value
CAL.P
Calibration password -See ‘USER
CALIBRATION’
0
to
99999
3
The following four parameters only appear when the correct password has been entered
CAL
Calibration type
Fact
Factory calibration
restored
FAct
USEr
User calibration enabled
The following three parameters are only shown if
<
USEr
>
is selected as the calibration type
Pnt.L
Start point low calibration
OFF
Calibration complete
OFF
on
Start low point calibration
Pnt.H
Start point high calibration
OFF
Calibration complete
OFF
on
Start high point
calibration
tArE.v
Tare
Value
This allows a fixed offset to be
applied to the displayed
reading. It must be set before
auto tare is started
-999.9
to
99999
display units
0.0
S.G
Specific
gravity
multiplier
For materials with specific
gravity different from water (1)
0.01
to
999.9
1.00
ScL.L
Scale Low
point
Defines the value which will be
displayed when the load is
removed from the cell
-999.9
to
99999
display units
0
ScL.H
Scale
High
point
Defines the value which will be
displayed when the load is
placed on the cell
-999.9
to
99999
display units
0
CAL1
CAL2
Part Number HA027240 Issue 5.0 Feb 2015
21