51
10.4 I/O
Pulses OUT
• Pulse quantity
• Pulses time ON
• Reverse flow rate
• Pulse output enable
Frequency output
• Full scale freq.
• Frequency output enable
Program. output
• Enabled/disabled
• Reverse Flow
• MAX flow th.
• Min flow th.
• MAX/MIN flow th.
• Batching
• Excitation failure
• Empty Pipe
Program. input
• Enabled/disabled
• Zeroing p+
• Zeroing p-
• Zeroing p+/p-
Batching
Progr. output logics
10.4.1 PULSE OUTPUT
• Pulse quantity
Select the volume of the pulses.
- MC608A (powered supply version)
Calculation of the parameters (values must be re
considered technical units of the converter)
[Vp=litres, Tp=seconds, Q=litres/seconds,
Qmax= ystem maximum flow]
Vp>Qmax*2Tp
Tp<Vp/2Qmax
WITH MC608B OR MC608R SELECT
A VOLUME LARGE ENOUGH SO THAT
DURING THE MAXIMUM FLOW OF
THE PROCESS THE TIME INTERVAL
BETWEEN TWO CONSECUTIVE
PULSES IS GREATER THAN 1 SECOND
IN ORDER TO PRESERVE THE
BATTERY LIFE.
CAUTION!
AVOID PULSES SATURATION, THIS
MAY LEAD TO AN ACCUMULATION
OF PULSES WHICH SHOULD
NOT BE PROVIDED IN REAL TIME
CONSIDERING THE
ACTUAL FLOW RATE.
NOTE
10. MENU DESCRIPTION