
50
EUROMAG | MC608 | TD210-1
www.euromag.com
10.3.4 ZERO FINDER
This setting is used to perform the device
zero calibration. Before performing the zero
calibration, ensure that:
1.
the sensor is full of liquid
2.
the liquid is perfectly stationary
3.
the sensor has correct electrical
grounding
10.3.5 FLOWRATE ALARMS
• MAX flow th.
Set the maximum value of flow rate in percentage
on the full scale value. This value is factory
disabled. Selectable range goes from +5%
threshold to 100% of the full scale value. To disable
the feature, select > 100%
• Min flow th.
Set the minimum value of flow rate in percentage
on the full scale value. This value is factory
disabled. Selectable range goes from 1% to
maximum threshold of 95% of the full scale value.
THE MC608 HAS AN AUTOMATIC ZERO
CALIBRATION FUNCTION. USE THE
ZERO FINDER ONLY WHEN STRICTLY
NECESSARY
NOTE
WHEN MAX OR MIN FLOW RATES ARE
ENABLED, A SYMBOL WILL APPEAR
ON THE DISPLAY IN CASE THESE
VALUES WILL BE REACHED.
NOTE
TO TEST THE EMPTY PIPE
DETECTION, PRESS THE
“TEST”BUTTON
NOTE
10.3.6 EMPTY PIPE THRESHOLD
(Available with sensors having 4
electrodes)
Allows you to adjust the sensitivity of the empty
pipe detection. Move the selection to the “E” if the
empty pipe alarm does not appear, to “F” if the
alarm does not disappear with the pipe full.
Summary of Contents for MC608 Series
Page 2: ......
Page 33: ...MC 608A MC 608B 33 Separate version 9 CONVERTER MC 608...
Page 49: ...fig_1033_1 49 10 MENU DESCRIPTION...
Page 66: ...66 EUROMAG MC608 TD210 1 www euromag com Note...
Page 67: ...67 NOTE...