2ad89664-e5f4-4907-bab1-9f7c76e7d9ef / DP10.0 / 2016-07-04
Page 37 of 43
Eurolyser Diagnostica GmbH
EXTERNAL INTERFACE DESCRIPTION
Serial Interface
Interface signals
Pin
Signal
1
2
TxD
3
RxD
4
5
GND
6
7
8
9
Interface parameters
Parameter
Value
Baud rate
9600
Data bits
8
Parity
None
Stop bits
1
Data format
Data is transmitted in blocks of data sets.
One data set contains the data from one analysis.
Data sets consist of data fields.
A data field consists of an identifier (7 characters) and its respective value or text.
Data fields are concluded with one carriage return and line feed.
Data sets are concluded with three carriage returns and line feeds.
Data Fields
Identifier
Format
Example
Remarks
Name:
Text
Name: John Doe
ID:
Text
ID: 1234
Sex:
Male/Female/Child/Control
Sex: Male
Sample:
Serum/Blood
Sample:Blood
Optional
HCT:
Value
HCT: 40
Optional
Testname
:
Value and unit
CRP: 2.48 mg/dl
Calculatedname:
Value and unit
eAG: 100 mg/dl
Optional
Range:
Value – Value
Range: 0.00 – 1.00
Optional
Time:
hh:mm
Time: 14:44
Date:
dd-mm-yyyy
Date: 08-02-2007
Operat:
Text
Operat:Max Muster
Optional
Lot No:
Value
Lot No:4111
Optional
Ser No:
Value
Ser No:Ba00001
Optional
USB Interface
The USB interface emulates the serial interface.