19
increases. When the bag size reaches approximately 14" the bag and system
capacitance overwhelms the discharge capacitance resulting in no detectable
difference between good and poor shielding material.
1. Set the FUNCTION select switch to MAN
'
V 400k or MAN
'
V 1.5k
2. Set the oscilloscope vertical sensitivity of both CH-1 and CH-2 to 50V/div, time
base to 20
P
sec and trigger level to 50V. Trigger mode should be set to CH-1
NORMAL. Invert CH-2. Both channels zero line should be superimposed on
the center line. (The trigger mode must first be set to AUTO to align the traces
then return to NORMAL.)
3. Turn on the HV
4. Initiate a discharge by either depressing the DISCHG switch on the front panel
or by activating the foot switch.
'
V testing can only be performed in the
Manual mode. Observe the waveform. It should look similar to that shown in
Figure 6.0-5. If the signal is off scale change the vertical sensitivity of both
channels and repeat the measurement.
Figure 6.0-5: Differential voltage set up waveform
5. Set the oscilloscope to MATH. Initiate a discharge. The differential pulse
detected by the capacitive sensor will be displayed on the scope along with
the actual differential voltage value.
'
V of less than 30 Volts is required for
MIL PRF 81705D material and is generally the level accepted by industry for
good performing shielding material.
6.0 SOFTWARE
The ETS Test Suite Manager is preloaded in the supplied computer (desktop or
laptop). However, the software is also provided on disc with the System. The supplied
computer does not require a password. If a password is desired, install using standard
computer password installation methods.
Make sure all cables are installed before opening the program.