![ETS 4431-EV Operating Manual Download Page 11](http://html1.mh-extra.com/html/ets/4431-ev/4431-ev_operating-manual_2434442011.webp)
10
5. USB to serial converter with 9-pin sub-D from computer to communications
cable (Model 4431-EV only)
NOTE: Do not sustitute
6. 9-pin sub-D from Converter to 25-pin sub-D to Model 4431-EV
7. Foot switch
8. IEC Line Cord (North American plug)
Figure 4.0-3b shows the jumpers and cables required to perform first time set up.
Figure 4.0-3b: Calibration cables and jumpers for first time set up
NOTE:
If the Model 4431 is used to replace a Model 431 discharge unit an adapter is
required to convert the 25-pin COMM PORT to a 3-pin DIN to receive its command
signals from the computer’s PRINTER Parallel Port (Contact ETS). If the computer has
only one output, it is necessary to install an A/B switch (supplied with early versions of
the ETS Energy Calculation Program) to operate a printer.
The Model 4431 utilizes a universal switching power supply and may be operated
directly from 90 - 260 VAC, 50/60 Hz line voltage (mains) with 0.5A resettable fuse that
is reset by powering down the unit, and after several seconds turning it back on.
The latest version of the ETS Energy Calculation Program is written for PCs running
Windows
£
2000/XP/VISTA/7. This program controls the Model 4431-EV, the Tektronix
TDS2022B oscilloscope used to capture the waveform and the computer to generate a
report containing all the data specified in ANSI/ESD STM11.31 plus the verification and
test data waveforms as shown in Figures 4.0-3 & 4.
5.0 SET-UP
Plug the AC line cord of the Model 4431 into a grounded outlet. This is a standard IEC
cord with North American grounded plug. For non-North American locations, a line cord
having the correct mains plug can be obtained from a local computer or electronics
store. Otherwise, cut off the plug and install the appropriate plug for the location.
5.1 ANSI/ESD STM11.31 Testing (Energy)
Refer to Figure 5.0-1 for the System set-up described below:
Cable to verify
discharge
waveform with
external CT-1
Shorting strap to
connect Dischg and
Capacitive
electrodes together
to measure 1kV
Dischg waveform