36
Operation
www.ets-lindgren.com
S
TATUS
M
ESSAGES OF THE
A
MPLIFIER IN
E
THERNET
R
EMOTE
C
ONTROL
M
ODE
The amplifier will return a status information if it is requested to do so by a data poll
command.
The status messages which are indicated on the front panel display can also be
transmitted as status information via the Ethernet remote control interface to the
control computer.
The status messages indicated on the front panel are encoded in the status text
according to the following table:
Query
Status message (String)
Meaning
*IDN?
ETS, 8100-041, xxx
Instrument identification: Manu-
facturer, Model number,
Serial number
AMP?
AMP_ON
Amplifier is set to operate mode
AMP_OFF
Amplifier is set to standby mode
CONTROL? LOCAL
Amplifier is in local control mode
GPIB
Amplifier is in GPIB control mode
LAN
Amplifier is in LAN control mode
TTL
Amplifier is in TTL control mode
STATUS?
SYSTEM_OK
Amplifier is working normally
Interlock External
External Interlock fault
Temperature X
Over temperature fault at sensor X
PowerSupply X
Power supply fault at output X
Control Fault
Internal control fault
Figure 15: Ethernet interface status messages
Summary of Contents for 8100-040
Page 1: ...Model 8100 041 RF Power Amplifier User Manual ...
Page 12: ...12 Maintenance www ets lindgren com This page intentionally left blank ...
Page 14: ...14 Specifications www ets lindgren com This page intentionally left blank ...
Page 16: ...16 Before You Begin www ets lindgren com This page intentionally left blank ...
Page 38: ...38 Operation www ets lindgren com This page intentionally left blank ...
Page 42: ...42 Warranty www ets lindgren com This page intentionally left blank ...
Page 44: ...44 EC Declaration of Conformity www ets lindgren com This page intentionally left blank ...