eSelect B2/B3 

Summary of Contents for eSelect B2

Page 1: ...eSelect B2 B3 ...

Page 2: ......

Page 3: ...ESPANOL 3 ...

Page 4: ...ails provides for punitive actions in case of illegal disposal of such waste FR DIRECTIVE RAEE 2002 96 CE ET MODIFICATION SUCCESSIVE 2003 108 CE CONCERNANT LES REBUTS D APPAREILLAGES ÉLECTRIQUES ET ÉLECTRONIQUES Le symbole ci dessous indique que le produit ne pas être éliminé comme un normal déchet urbain Les Appareillages Électriques et Électroniques AEE peuvent contenir des matériaux nocifs pour...

Page 5: ...ES PH RX CL 14 7 3 PROGRAMMING PROCEDURE 14 7 3 1 SIMPLE MENU MODE 14 7 3 2 EXPERT MENU MODE 15 7 4 PROGRAMMING MENU SETUP 16 7 4 1 MENU MODE 16 7 4 2 MENU METER TYPE 17 7 4 3 T SENSOR 17 7 4 4 SET TEMPERATURE 18 7 4 5 CALIBRATION MENU TIMEOUT 18 7 4 6 DELAY AT STARTUP 18 7 4 7 AUX OUTPUT MODE 19 7 4 8 YEAR MONTH DAY HOUR MINUTE 20 7 4 9 SET PASSWORD 20 7 4 10 SET DEFAULTS 20 7 5 MENU SETPOINT1 SE...

Page 6: ...resolution Chlorine range 0 2 0 20 0 200 0 2000 ppm 0 001 0 01 0 1 1 ppm Temperature range 0 100 C 0 1 C Level control PT100 connection Relay output 6A resistive load 1A inductive load 1 GENERAL INFORMATIONS Electronic instruments controlling electrochemical parameters such as pH Redox or Chlorine are widely used in swimming pools waterworks and water treatment plants The eSelect Series Controller...

Page 7: ...3 3 CONNECTIONS 3 1 eSelect B2 POWER SUPPLY CARD CONNECTIONS ...

Page 8: ...4 3 2 eSelect B2 TTL OUTPUT CONNECTIONS ...

Page 9: ...5 3 3 eSelect B3 POWER SUPPLY CARD CONNECTIONS ...

Page 10: ...6 3 4 eSelect B3 TTL OUTPUT CONNECTIONS ...

Page 11: ...necting pH or Redox probes plus a 4 pin connector for connecting a Chlorine probe are provided in the bottom of the instrument The connection diagram is as follows Power supply 4 pins connector Pin 1 White 5V Pin 2 Not connected Pin 3 Not connected Pin 4 Brown 5V ...

Page 12: ...8 4 CONTROL PANEL DESCRIPTION 4 1 eSelect B2 CONTROL PANEL 4 2 eSelect B3 CONTROL PANEL ...

Page 13: ...ing menus E g when selecting the type of measurement allows the user to shift between pH Rx and Chlorine Plus symbol Increases numbers and defines functions within specific programming menus E g when selecting the type of measurement allows the user to shift between pH Rx and Chlorine Shift right Used to select the digit to modify when setting passwords Used to change Language also OK Allows the u...

Page 14: ...10 5 DIMENSIONS ...

Page 15: ... mA output 1 and 2 measurement at 4 mA 0 00 0 0 000 9 mA output 1 and 2 measurement at 20 mA 14 00 1000 1 000 10 Alarm high threshold 14 00 1000 10 11 Alarm low threshold 0 00 0 0 000 12 Alarm overdosage OVER 99 59 h m 99 59 h m 99 59 h m 13 Menu mode SIMPLE SIMPLE SIMPLE 14 Password OFF OFF OFF 15 Temperature unit C C C 16 Temperature compensation mode Manual 25 C Manual 25 C Manual 25 C 17 Calib...

Page 16: ...nts pH Rx or Chlorine To perform that when switching on the controller for the first time the operator have to select the intended type of measurement by operating as follows Upon switching on the instrument press OK key to enter in the MENU SETUP Than press OK to enter in MENU MODE and change SIMPLE with EXPERT by press or key Than change the METER TYPE in the complete MENU SETUP Press OK key to ...

Page 17: ...TION MENU TIMEOUT 05 00 DELAY AT STARTUP 00 05 AUX OUTPUT MODE OFF YEAR 2013 MONTH 5 DAY 15 MENU SETUP 1 2 MENU MODE EXPERT METER TYPE pH T SENSOR NONE SET TEMPERATURE 25 0 CALIBRATION MENU TIMEOUT 05 00 DELAY AT STARTUP 00 05 AUX OUTPUT MODE OFF YEAR 2013 MONTH 5 DAY 15 After choose the METER TYPE press ESC key until see the main screen Please note the ESC key permits to return to previous screen...

Page 18: ...METER 2 for B2 controller and METER 1 METER 2 and METER 3 for B3 controller When setting the instrument it is advisable to follow the programming procedure outlined below Conversely if only a single parameter must be changed it is advisable to go directly to the menu comprising the concerned function and perform the necessary changes or settings Please note there are two different MENU MODE SIMPLE...

Page 19: ...its to set the following parameters by use MENU MODE SIMPLE or EXPERT METER TYPE pH Rx Cl 20ppm Cl 2ppm Cl 200ppm Cl 2000ppm TSENSOR NONE or PT100 SET TEMPERATURE put the current TEMPERATURE is PT100 is not setted CALIBRATION MENU TIMEOUT set the TIMEOUT time for CALIBRATION procedure DELAY STARTUP set the startup DELAY TIME for the instrument AUX OUTPUT MODE set OFF or CLEANING or FLOCCULANT YEAR...

Page 20: ... TIMEOUT 05 00 DELAY AT STARTUP 00 05 AUX OUTPUT MODE OFF YEAR 2013 MONTH 5 DAY 15 7 4 PROGRAMMING MENU SETUP 7 4 1 MENU MODE Starting from the MAIN SCREEN press OK key until enter in MENU MODE Than change SIMPLE to EXPERT by pressing key or key Change SIMPLE in EXPERT by press or key Chose the parameter by press or key Press OK key to confirm the choice and set the value for the parameter on righ...

Page 21: ...AT STARTUP 00 05 AUX OUTPUT MODE OFF YEAR 2013 MONTH 5 DAY 15 MENU SETUP 1 2 MENU MODE EXPERT METER TYPE pH T SENSOR NONE SET TEMPERATURE 25 CALIBRATION MENU TIMEOUT 05 00 DELAY AT STARTUP 00 05 AUX OUTPUT MODE OFF YEAR 2013 MONTH 5 DAY 15 7 4 2 MENU METER TYPE Starting from the MENU SETUP screen enter in METER TYPE by pressing key or key After selected METER TYPE press OK key to set the meter typ...

Page 22: ...e to set the real temperature as show in next paragraph 7 4 4 SET TEMPERATURE Starting from the MENU SETUP screen if the PT100 sensor is not setted enter in SET TEMPERATURE menu by pressing key or key After selected SET TEMPERATURE press OK key to set the temperature value 7 4 5 CALIBRATION MENU TIMEOUT It represent the exit delay time from the programming menu in case no keys are pressed during t...

Page 23: ...screen enter in AUX OUTPUT MODE by pressing key or key After selected AUX OUTPUT MODE press OK key to set between OFF CLEANING and FLOCCULANT Once one of the two functions is activated flocculant or cleaning it is necessary to set the auxiliary outputs that need to be activated by select the days in which the auxiliary output needs to be activated 7 4 7 1 CLEANING AUX OUTPUT MODE Starting from the...

Page 24: ...and the relating 6 digits code The password can be any number between 000000 and 999999 To set the value for any digit use or key and move to the next digit by press key After setting all 6 digit press OK key to set the next parameter 7 4 10 SET DEFAULTS Starting from the MENU SETUP screen enter in SET DEFAULT by pressing key or key Than press OK key and set one of the followings NO to don t reset...

Page 25: ... key to move the selector on MENU SETPOINT1 or MENU SETPOINT2 Press OK key to enter Please note starting from SIMPLE MENU MODE into the SETPOINT1 and SETPOINT2 it is possible to set only the SETPOINT value and the working MODE acid or alkaline Press OK key to enter into MENU MAIN SCREEN Chose the parameter by press or keys Press OK key to confirm the choice and set the value for the parameter on r...

Page 26: ...s the action tends to decrease or increase the measurement After set the MODE parameter press OK key to move the cursor on right column and set the value by press or keys Press OK key to confirm the setted value 7 5 3 HYSTERESIS After defining the type of actuation setting the hysteresis value is the next step After set the HYSTERESIS parameter press OK key to move the cursor on right column and s...

Page 27: ...L output is ON After set the PULSE WIDTH parameter press OK key to set the value 7 5 5 2 VAL MAX F This parameter represent the measure where the frequency of the TTL output is on MAX value After set the VAL MAX F parameter press OK key to set the value 7 5 5 3 MAX FREQUENCY This parameter represent the maximum frequency of the TTL output After set the MAX FREQUENCY parameter press OK key to set t...

Page 28: ...s key to move the selector on MENU CALIBRATION Press OK key to enter Press OK key to enter into MENU MAIN SCREEN Chose the parameter by press or keys Press OK key to confirm the choice and set the value for the parameter on right column Modify the value for the parameter by press or and keys Press OK key to confirm the choice and return on left column to set an other parameter Calibration Menu for...

Page 29: ...ution and press OK on CALIBRATE BUFFER 4 9 When the reading is stable adjust the value by using the and keys and press key to confirm The instrument is so calibrated For Redox and Chlorine calibrate first calibration point by using buffer solution e g 650mV o by using a photometer CALIBRATE BUFFER for REDOX and CALIBRATE DPD1 for CLORINE Second calibration point available in EXPERT mode only is th...

Page 30: ... ESELECT B3 The proximity sensor when is inserted in the probe holder signal the presence of water in the installation and therefore the need to start the inspection In order to activate the controller the proximity sensor should be Normally Closed When there is no water flow the contact on REMOTE INPUT is open than a STAND BY alarm will shows on the display It is possible to use one proximity sen...

Page 31: ... To facilitate the priming of the pumps it is possible to manually activate the output of the SETPOINT Such procedure is accessible even in presence of password by pressing simultaneously two keys as following described while the visualization of the measure is in progress The procedure of priming of the pumps involves Activation of the relay SETPOINT 4 20mA output is set to 20mA If in ON OFF mode...

Page 32: ......

Page 33: ......

Page 34: ......

Page 35: ...COD DMU00193ML1 A 09 2013 ...