N.B.:
Only necessary when changing the calibration factor
-
For precise calibration, it is essential that the unit is free of air during the measuring
process.
-
Let the pump run for around 1 minute first.
-
If possible, use an outlet pipe with a drip-stop valve.
-
Use a calibrated measuring vessel with as large a capacity as possible.
-
Follow the steps in the description exactly.
1
Press
TOTAL
then
press
RESET
The display will show
0,00
2
Press
FLOW RATE
+
SELECT
The display will show
for example
9 1 0 0 4 5
now convert this figure as follows
Example:
9 1 0 0 4 5
1
9
first number e.g. 9 9 x 1 =
9
5
5
second number e.g. 1 1 x 5 =
5
12
0 0 x 12=0
1
0
3
12
9
45
Add them up
71
3
Press
TOTAL
+
FLOW RATE
The display will show
Cd 0
4
The CODE number calculated earlier can now
be entered by pressing the
SELECT
key
Example
71
.