1. P
REPARATION
FOR
DIOPTER
COMPENSATION
AND
IPD
ADJUSTMENT
Use the focusing test bar
The bar is a standard accessory for accurate adjustment of the microscope.
Insert it into the correct poison of slit lamp. With the focal plane facing to the objective lens.
Remove the bar after testing.
1. Test bar
2. Focus plane
Brightness adjustment
Switch ON the main power.
Rotate the light dimmer to the central position.
Set the slit width at 2~3 mm.
With:
•
- : Darker
•
+ : Brighter
1
1
2
3
SL300L / SL400L - Slit lamp microscope > V2 - 03-2018
28
U
SER
MANUAL
> V. O
PERATION
PROCEDURES