50
12-2
Replacement of the optical control board ASSY
1)
Disassemble the main unit cover L and R (refer to 12-1).
2)
Remove 6 harnesses that are connected to the optical control board ASSY.
Optical unit cover ASSY
Harnesses
3)
Unscrew 3 fixing screws of the optical control board ASSY and remove the ASSY.
This is the end of disassembling process. Assemble the control board with the backward
process of disassembly.
Fixing screws