Cytotoxic Lead Shielded Class II Biological Safety Cabinets
Decontamination
Decontamination may frequently be carried out by means of formaldehyde fumigation or using other
decontamination agents, such as chlorine dioxide or hydrogen peroxide. Decontamination process should only
be carried out by trained personnel.
In any of the following eventualities, the user should ensure that the BSC has been properly decontaminated,
keeping in mind the nature of the pathogens used:
At the time of moving/relocating the BSC
At the time of changing the type of work being carried out in the BSC
Before accessing contaminated areas for servicing, for example - filter replacement