Page 23 of 34
6.0
CALIBRATION
&
DIAGNOSTIC
PROCEDURES
SuiteSentry Single Screener control functions are set at the time of product installation by the
installer.
CAUTION! Changing of the control settings should be
approached with care as they will alter the behavior of the unit.
Switch 2 -- Activates
AUDIO ALARM
Down is
OFF
- Up is
ON
SENSITIVITY CONTROL
Adjusts the ferromagnetic
disturbance threshold
required to trigger an
alarm.
Sensitivity increases
with clockwise rotation.
Sensitivity control shown
in the initial (detent)
position.
Switch 1 -- Activates PIR
motion detector
Down is
OFF
- Up is
ON
RS 232 Connector -
Service Port
V2 -- Audio-alarm volume
control.
Clockwise rotation
decreases the alarm
volume.
V1 -- PIR-Motion detector
sensitivity control.
Clockwise rotation decreases
the alarm activation footprint.
Figure 14: SuiteSentry Single Screener Controls
Control Functions - Top Control Panel
(under slide)