38
Specifications
ets-lindgren.com
HI-6153
S
PECIFICATIONS
Dynamic Range:
2.0
–800 Volts per meter (V/m)
Ranges:
Single
Typical Frequency Response:
10 MHz
–40 GHz
10 MHz
–100 MHz: +3 dB, -4 dB
100 MHz
–1 GHz: +3 dB, -0.5 dB
1 GHz
–18 GHz: +4 dB, -2 dB
18 GHz
–40 GHz: +3.5 dB, -4.5 dB
Typical Frequency Response
with Correction:
10 MHz
–18 GHz: ± 0.9 dB
18 GHz
–40 GHz: ± 1.1 dB
Linearity:
±0.5 dB
Isotropicity:
±1.0 dB < 18 GHz
Overload Withstand:
1,500 V/m
Physical Interface:
Duplex optical fiber
(62.5 micron multimode)
FC
connectors for laser cable,
integral 1-m optical cable
ST
connector for transmitter cable,
integral 1-m optical cable
Probe Mount:
1/4
–20 UNC tapped hole (internal thread)