Maintenance
ets-lindgren.com
If the probe shields need to be replaced, you may order them from
ETS-Lindgren. Follow these steps to replace the probe shields:
1.
Remove the nylon screws from the probe shield.
2.
Lift the shield straight up and away from the housing.
Note:
Lift the shield high enough to clear the internal antenna
assembly, or you may damage the assembly. Never touch the
antenna assembly. You may transfer a static charge to the
assembly that may distort or disable the measuring capability in that
axis.
3.
With the probe in a protective, supportive position, carefully replace the
probe shield by lowering it straight down onto the housing.
Note:
Make sure to clear the internal antenna assembly when
lowering the shield or you may damage the assembly.
4.
Replace the screws snugly without over-tightening them.
Note:
Over-tightening the screws may damage the screw heads. If
you require additional nylon screws, you may order them from
ETS-Lindgren or obtain them locally. Non-conductive screws must
be used for this application. Never use metal screws.
5.
Repeat this procedure for the other shields, if necessary.
Summary of Contents for ETS-Lindgren HI-6006
Page 1: ...HI 6122 HI 6105 HI 6153 HI 6023 HI 6006 HI 6053 EMC Field Probes User Manual ...
Page 10: ...x ets lindgren com This page intentionally left blank ...
Page 20: ...20 Typical Configurations ets lindgren com This page intentionally left blank ...
Page 34: ...34 Maintenance ets lindgren com This page intentionally left blank ...
Page 44: ...44 Specifications ets lindgren com This page intentionally left blank ...
Page 72: ...72 Warranty ets lindgren com This page intentionally left blank ...
Page 138: ...138 ETSProbe DLL User Guide ets lindgren com This page intentionally left blank ...