Esam E1200 User Manual Download Page 10

 

MANUALE D’USO / USER’S GUIDE

 

 



E1200 

[email protected]     www.esam.biz 

page 10 - 16 

rev. 1.07 

0714 

 

 

Register 

Type 

Read / Write 

Label 

Description 

Aron 

Unit 

234-235 

Float 

Read only 

PMN 

Potenza attiva negativa media / 

Average Negative Active Power

 

 

236-237 

Float 

Read only 

QMP 

Potenza reattiva positiva media / 

Average Positive Reactive Power

 

 

VAR 

238-239 

Float 

Read only 

QMN 

Potenza reattiva negativa media / 

Average Negative Reactive Power

 

 

VAR 

240-241 

Float 

Read only 

PMP1 

Potenza att. pos. media fase 1 / 

Phase 1 Average Pos. Act. Power

 

 

242-243 

Float 

Read only 

PMN1 

Potenza att. neg. media fase 1 / 

Phase 1 Average Neg. Act. Power

 

 

244-245 

Float 

Read only 

QMP1 

Potenza reatt. pos. media fase 1 / 

Phase 1 Average Pos. React. Power

 

 

VAR 

246-247 

Float 

Read only 

QMN1 

Potenza reatt. neg. media fase 1 / 

Phase 1 Average Neg. React. Power

 

 

VAR 

248-249 

Float 

Read only 

PMP2 

Potenza att. pos. media fase 2 / 

Phase 2 Average Pos. Act. Power

 

 

250-251 

Float 

Read only 

PMN2 

Potenza att. neg. media fase 2 / 

Phase 2 Average Neg. Act. Power

 

 

252-253 

Float 

Read only 

QMP2 

Potenza reatt. pos. media fase 2 / 

Phase 2 Average Pos. React. Power

 

 

VAR 

254-255 

Float 

Read only 

QMN2 

Potenza reatt. neg. media fase 2 / 

Phase 2 Average Neg. React. Power

 

 

VAR 

256-257 

Float 

Read only 

PMP3 

Potenza att. pos. media fase 3 / 

Phase 3 Average Pos. Act. Power

 

 

258-259 

Float 

Read only 

PMN3 

Potenza att. neg. media fase 3 / 

Phase 3 Average Neg. Act. Power

 

 

260-261 

Float 

Read only 

QMP3 

Potenza reatt. pos. media fase 3 / 

Phase 3 Average Pos. React. Power

 

 

VAR 

262-263 

Float 

Read only 

QMN3 

Potenza reatt. neg. media fase 3 / 

Phase 3 Average Neg. React. Power

 

 

VAR 

264-265 

Float 

Read only 

PMMP 

Max Potenza media att. pos. / Max 

Average Positive Active Power

 

 

266-267 

Float 

Read only 

PMMN 

Max Potenza media att. neg. / Max 

Average Negative Active Power

 

 

268-269 

Float 

Read only 

QMMP 

Max Potenza media reatt. pos. / Max 

Average Positive Reactive Power

 

 

VAR 

270-271 

Float 

Read only 

QMMN 

Max Potenza reatt. neg. media / Max 

Average Negative Reactive Power

 

 

VAR 

272-273 

Float 

Read only 

PMMP1 

Max Potenza att. pos. media fase 1 / 

Max Phase 1 Average Pos. Act. Power

 

 

274-275 

Float 

Read only 

PMMN1 

Max Potenza att. neg. media fase 1 / 

Max Phase 1 Average Neg. Act. Power

 

 

276-277 

Float 

Read only 

QMMP1 

Max Potenza reatt. pos. media fase 1 / 

Max Phase 1 Average Pos. React. Power

 

 

VAR 

278-279 

Float 

Read only 

QMMN1 

Max Potenza reatt. neg. media fase 1 / 

Max Phase 1 Average Neg. React. Power

 

 

VAR 

280-281 

Float 

Read only 

PMMP2 

Max Potenza att. pos. media fase 2 / 

Max Phase 2 Average Pos. Act. Power

 

 

282-283 

Float 

Read only 

PMMN2 

Max Potenza att. neg. media fase 2 / 

Max Phase 2 Average Neg. Act. Power

 

 

284-285 

Float 

Read only 

QMMP2 

Max Potenza reatt. pos. media fase 2 / 

Max Phase 2 Average Pos. React. Power

 

 

VAR 

286-287 

Float 

Read only 

QMMN2 

Max Potenza reatt. neg. media fase 2 / 

Max Phase 2 Average Neg. React. Power

 

 

VAR 

288-289 

Float 

Read only 

PMMP3 

Max Potenza att. pos. media fase 3 / 

Max Phase 3 Average Pos. Act. Power

 

 

290-291 

Float 

Read only 

PMMN3 

Max Potenza att. neg. media fase 3 / 

Max Phase 3 Average Neg. Act. Power

 

 

292-293 

Float 

Read only 

QMMP3 

Max Potenza reatt. pos. media fase 3 / 

Max Phase 3 Average Pos. React. Power

 

 

VAR 

294-295 

Float 

Read only 

QMMN3 

Max Potenza reatt. neg. media fase 3 / 

Max Phase 3 Average Neg. React. Power

 

 

VAR 

 

 

 

 

 

 

 

1000

1127 

Int 

Read only 

 

Ampiezza armoniche V1 / 

Amplitude of V1 harmonics

 

 

(

3

)

 

1128

1255 

Int 

Read only 

 

Ampiezza armoniche I1 / 

Amplitude of I1 harmonics

 

 

(

3

)

 

1256

1383 

Int 

Read only 

 

Ampiezza armoniche V2 / 

Amplitude of V2 harmonics

 

 

(

3

)

 

1384

1511 

Int 

Read only 

 

Ampiezza armoniche I1 / 

Amplitude of I2 harmonics

 

 

(

3

)

 

1512

1639 

Int 

Read only 

 

Ampiezza armoniche V3 / 

Amplitude of V3 harmonics

 

 

(

3

)

 

1640

1767 

Int 

Read only 

 

Ampiezza armoniche I3 / 

Amplitude of I3 harmonics

 

 

(

3

)

 

 

 

 

 

 

 

 

2100

2149 

- - - 

Read only 

 

Registri blocco virtuale / 

Virtual register block

 

 

 

 
 

Tab. 2:  REGISTRI DEI MINIMI E MASSIMI DELLE MISURE / 

REGISTERS OF MIN. AND MAX. MEASURED VALUES

 

 

Min Reg. 

Max reg. 

Type  Read/Write 

Description 

Aron  Unit 

2200-2201  2400-2401 

Float 

Read only 

Tensione fase 1 / 

Voltage Phase 1

 

 

2202-2203  2402-2403 

Float 

Read only 

Corrente fase 1 /

 Current Phase 1

 

 

2204-2205  2404-2405 

Float 

Read only 

Potenza attiva fase 1 / 

Active power Phase 1

 

 

2206-2207  2406-2407 

Float 

Read only 

Potenza reattiva fase 1 / 

Reactive power Phase 1

 

 

VAR 

2208-2209  2408-2409 

Float 

Read only 

Tensione concatenata fase 1-2 / 

Linked Voltage Phase 1-2

 

 

2210-2211  2410-2411 

Float 

Read only 

Valore picco V1 nell’ultimo periodo AC / 

V1 peak value in last AC period

 

 

2212-2213  2412-2413 

Float 

Read only 

Valore picco I1 nell’ultimo periodo AC / 

I1 peak value in last AC period

 

 

2214-2215  2414-2415 

Float 

Read only 

Potenza Apparente fase 1 / 

Apparent power Phase 1

 

 

VA 

2216-2217  2416-2417 

Float 

Read only 

Fattore di potenza fase 1 / 

Power factor Phase 1

 

 

 

2218-2219  2418-2419 

Float 

Read only 

Distorsione armonica totale V1 / 

Total harmonic distortion V1

 

 

2220-2221  2420-2421 

Float 

Read only 

Distorsione armonica totale I1 / 

Total harmonic distortion I1

 

 

2222-2223  2422-2423 

Float 

Read only 

Tensione fase 2 / 

Voltage Phase 2

 

 

2224-2225  2424-2425 

Float 

Read only 

Corrente fase 2 /

 Current Phase 2

 

 

2226-2227  2426-2427 

Float 

Read only 

Potenza attiva fase 2 / 

Active power Phase 2

 

 

2228-2229  2428-2429 

Float 

Read only 

Potenza reattiva fase 2 / 

Reactive power Phase 2

 

 

VAR 

2230-2231  2430-2431 

Float 

Read only 

Tensione concatenata fase 2-3 / 

Linked Voltage Phase 2-3

 

 

Summary of Contents for E1200

Page 1: ...ccia Ethernet configurazione dello strumento e visualizzazione misure da browser Web acquisizione dati da sistemi Scada con protocollo Modbus TCP inserzione trifase con 3 TA o 2 TA ARON o monofase con 1 TA rapporti TA e TV configurabili montaggio su profilato DIN EN50022 35 The Network analyzer E1200 entirely designed and developed by ESAM is built to comply with all the modern requirements of mea...

Page 2: ...attiva uscente Active Energy Out Ea Ea1 Ea2 Ea3 Wh Energia reattiva entrante Reactive Energy In Er Er1 Er2 Er3 VARh Energia reattiva uscente Reactive Energy Out Er Er1 Er2 Er3 VARh Potenza attiva media entrante ultimi 15 minuti Average Active Power In last 15 minutes Pm Pm1 Pm2 Pm3 W Potenza attiva media uscente ultimi 15 minuti Average Active Power Out last 15 minutes Pm Pm1 Pm2 Pm3 W Potenza rea...

Page 3: ...hase input voltage range 0 1 2Vn nominal input current 5A input current range 0 1 2 In continuous overload 1 5Vn 2In instantaneous overload 1 sec 2Vn 20In programmable external TV ratio 1 99999 programmable external TV ratio 1 99999 averaging time interval AC period 1 sec accuracy voltage current active power 0 2 of reading 5 100 of input range 45 65 Hz temperature coefficient 0 01 C RS485 serial ...

Page 4: ...ia di rete Ethernet a 10 Mbit s in reti miste 10 100 1000 Mbit s ciò non costituisce un problema purché l E1200 venga collegato a uno switch di rete Evitare invece di usare uno hub in quanto ciò rallenterebbe anche i dispositivi in grado di comunicare fra di loro a velocità maggiore Lo strumento è dotato di un connettore di rete standard RJ45 e ha la seguente configurazione di default Indirizzo IP...

Page 5: ...legare l E1200 a uno switch di rete con un cavo normale se l E1200 è stato correttamente configurato a questo punto sarà accessibile con un browser da qualsiasi PC collegato alla rete Lo strumento è completamente configurabile per mezzo del web server Il web server consente inoltre la visualizzazione di tutti i valori misurati e calcolati The E1200 network analyzer does not support DHCP and theref...

Page 6: ...put modules V mA Outputs For each E3000 analog output max 4 E3000 modules 12 analog outputs Output type 0 10V 4 20mA Normal inverted output e g 4 20 or 20 4 mA Output function for power factor readings Low range of output variable High range of output variable Reset readings Individual commands allow to reset Max average power values Min max values of all the readings Energy values SERVER MODBUS T...

Page 7: ...pronti a ricevere quando l E1200 inizia a rispondere SWFP ordine di trasmissione delle due word dei valori floating point valori validi 0 1 vedi protocollo Modbus nota 2 Parametro configurabile solo per la seconda porta seriale PROT selezione Modbus slave PROT 2 default oppure Modbus master PROT 4 per pilotaggio moduli di output analogico E3000 Impostazioni default di fabbrica per entrambe le port...

Page 8: ...23 bits SWFP Command Setting to 0 word A will be sent before word B setting 1 word B will be sent before word A Modbus RTU implemented exception codes 1 illegal function invalid or non implemented function 2 illegal data address read or write request at an invalid address Ex Non existent register read one word only of a floating point value 3 illegal data value write request of an invalid value e ...

Page 9: ... picco I3 nell ultimo periodo AC I3 peak value in last AC period A 158 159 Float Read only S3 Potenza Apparente fase 2 Apparent power Phase 2 VA 160 161 Float Read only PF3 Fattore di potenza fase 3 Power factor Phase 3 162 163 Float Read only THD V3 Distorsione armonica totale V3 Total harmonic distortion V3 164 165 Float Read only THD I3 Distorsione armonica totale I3 Total harmonic distortion I...

Page 10: ...Phase 2 Average Neg Act Power W 284 285 Float Read only QMMP2 Max Potenza reatt pos media fase 2 Max Phase 2 Average Pos React Power VAR 286 287 Float Read only QMMN2 Max Potenza reatt neg media fase 2 Max Phase 2 Average Neg React Power VAR 288 289 Float Read only PMMP3 Max Potenza att pos media fase 3 Max Phase 3 Average Pos Act Power W 290 291 Float Read only PMMN3 Max Potenza att neg media fas...

Page 11: ... potenza del sistema trifase The three phase power factor 2280 2281 2480 2481 Float Read only Corrente di neutro Neutral current A Tab 3 REGISTRI DEI PARAMETRI PARAMETER REGISTERS Register Type Read Write Label Description Value Unit 300 301 Float Read only TAP Ingresso nominale I Rated I Input A 302 303 Float Read only TVP Ingresso nominale V Rated V Input V 304 305 Float Read Write ETAR Rapporto...

Page 12: ...ll uscita function applied to output Tab 8 2708 Int Read Write AOS13 E3000 mod 1 out 3 selezione misura in uscita output reading selection 6 100 180 2709 Int Read Write AOT13 E3000 mod 1 out 3 tipo di uscita output type Tab 7 2710 Int Read Write AOI13 E3000 mod 1 out 3 uscita normale invertita normal inverted output 0 1 2711 Int Read Write AOF13 E3000 mod 1 out 3 funzione applicata all uscita func...

Page 13: ... 2 out 2 fine campo misura output high range 7 2820 2821 float Read Write AOL23 E3000 mod 2 out 3 inizio campo uscita output low range 7 2822 2823 float Read Write AOH23 E3000 mod 2 out 3 fine campo misura output high range 7 2824 2825 float Read Write AOL31 E3000 mod 3 out 1 inizio campo uscita output low range 7 2826 2827 float Read Write AOH31 E3000 mod 3 out 1 fine campo misura output high ran...

Page 14: ... which word has to be sent first so set this parameter to have them sent in the order that your master Modbus equipment understands MSB LSB seeeeeee emmmmmmm mmmmmmmm mmmmmmmm word A word B s segno del numero 1 se il valore è negativo 0 se il valore è positivo Sign bit negative it is 1 positive it is 0 e esponente a 8 bit 8 bit exponent m mantissa del numero 23 bit The mantissa which is code in 23...

Page 15: ... campo verificare che la rete RS485 sia cablata a regola d arte soprattutto per collegamenti di lunghezza elevata e con molti slave connessi usare doppino schermato di buona qualità collegare la calza al terminale 0 degli slave mettere la calza a terra in un unico punto evitare diramazioni della linea e collegamenti a stella montare la resistenza di terminazione 120 ohm ai due estremi della linea ...

Page 16: ...ve cable too long or wrong type defective network switch 2 Check if the web server works otherwise verify the E1200 network configuration IP address netmask as detailed in the Ethernet interface section of this manual The E1200 IP address must be different from any already assigned static IP address and out of the range of dynamic IP addresses managed by DHCP 3 Check if the E1200 and your Modbus T...

Reviews: