5
GB
GB
GB
GB
GB
RTD INPUT
RTD INPUT
RTD INPUT
RTD INPUT
RTD INPUT
Fig. 7 RTD INPUT WIRING
NOTE
NOTE
NOTE
NOTE
NOTE:
1) Don’t run input wires together with power
cables.
2) Pay attention to the line resistance; an high
line resistance may cause measurement
errors.
3) When shielded cable is used, it should be
grounded at one side only to avoid ground loop
currents.
4) The resistance of the 3 wires must be the
same.
16
RTD
14
15
16
RTD
14
15
TC type J, K, L, N, T, E, U and PLAT.II
TC type J, K, L, N, T, E, U and PLAT.II
TC type J, K, L, N, T, E, U and PLAT.II
TC type J, K, L, N, T, E, U and PLAT.II
TC type J, K, L, N, T, E, U and PLAT.II
Fig 6.B INPUT WIRING FOR TC TYPE J, K, L, N,
T, E, U, Plat. II.
NOTE
NOTE
NOTE
NOTE
NOTE:
1) Do not run input wires with power cables.
2) For TC wiring use proper compensating cable
preferable shielded (see "Appendix B" of the
LHI or LHL user manual).
3) when a shielded cable is used, it should be
connected to one side only.
11
+
_
Shield
11
+
_
Shield
15
15
lhlr-1-a0.pmd
05/06/2006, 9.20