How to use this device
72/98
–
Only touch new lamps in the handling zone.
–
Test the safety switch to ensure that it works.
The safety switch ensures that the device switches off in the event of a
faulty filter panel or when a filter panel is missing.
–
Press the button and release it.
If the button springs back automatically then it is functional.
–
Replace the filter panel.
12.16.5
Cleaning and disinfection of the sensor
Not applicable for this device.
12.16.6
Cleaning or replacing lamps in the base
–
Lift up the acrylic panel and then remove the lamp covers and the
partition: see page 64.
Replacing lamps: see page 69.
–
Please maintain the correct order of the Genesis Type R HYBRID
Performance low pressure lamps and Genesis Type U HYBRID
Performance UV low pressure lamps.
a1: Genesis Type R HYBRID Performance low pressure lamps
a2: Genesis Type U HYBRID Performance UV low pressure lamps
a1
a2
19192 / 0