ergometer system ergoselect 50
Operating Mode
continuous operation
Power
100 ‑ 240 V / 50 ‑ 60 Hz
(60 VA max.)
specification power cord US:
SJT 2xAWG16 125 V / 10 A
„hospital“ or „ hospital grade“
specification internal backup battery:
IEC: CR 2032 / 3 V 230 mAh
Braking Principle
computer‑controlled eddy current brake
with torque measurement;
speed independent to DIN VDE 0750‑0238
Load Range
6 ‑ 400 watts, speed independent
(see diagrams)
Speed Range
30 to 130 RPM
Deviation of Measured Load
to DIN VDE 0750‑0238
Load Increments
user programmable
Internal Protocols
Control Terminal P:
• 5 fixed incremental exercise test protocols (e.g. WHO)
• 10 user‑programmable protocols
• manual load control
Permitted Patient Weight
160 kg
Saddle Height Adjustment
continuous, for patients between 120 cm and 210 cm
manual adjustment of saddle height
Handlebar Adjustment
for patient heights between 120 cm and 210 cm
continuous handlebar adjustment over 360°
Crank Length
170 mm (cranks with adjustable length are optional
accessories)
Displays
LCD:
68 x 34 mm, 128 x 64 pixels
additional LED display for speed (RPM)
Interfaces
PORT 1 (DSUB‑9‑pole):
digital remote control by PC or ECG recorder,
remote start of ECG recorder (optional)
USB:
digital remote control by PC (driver required)