Maintenance and servicing
36
/ 46
801
16-
443
10.
16
Operation suction
Test setup
•
As for the previous test.
Test procedure
Operation suction without
run-on time
1. Set IES 2 to:
– Operation suction: 80%
– Basic suction: 2%
– Run-on time: 0 sec
2. Keep footswitch pedal pressed. The suction must remain activated as
long as the pedal is kept pressed.
3. Release the footswitch pedal. The suction must be deactivated imme-
diately.
Operation suction with
run-on time
1. Set IES 2 to:
– Operation suction: 80%
– Basic suction: 2%
– Run-on time: 5 sec
2. Keep footswitch pedal pressed. The suction must remain activated as
long as the pedal is kept pressed.
3. Release the footswitch pedal. The suction must be deactivated after
approx. 5 sec.
Blocked suction detection
Test setup
•
As for the previous test.
Test procedure
1. Ascertain and note the sensitivity of blocked suction detection in the
Setting mode (P. 3), see page 26. Set the blocked suction detection to
7 for this test.
2. Exit the Setting mode. Switch unit off and back on for this purpose.
3. Keep footswitch pedal pressed. The suction must be activated.
4. Keep the main filter input covered up (to simulate blocked suction).
Blocked suction detection works if:
– the
System blocked / Occlusion
display lights red
– a warning signal sounds (1 sec on, 1 sec off)
– the suction is switched off immediately (without run-over time).
5. The sensitivity of blocked suction detection is reset to the old value.
Summary of Contents for IES 2
Page 1: ...10 16 80116 443 SERVICE MANUAL IES 2 1 3 x...
Page 2: ......
Page 3: ...SERVICE MANUAL IES 2...
Page 18: ...4 Installation 18 46 80116 443 10 16...
Page 19: ...19 46 5 Circuit Descriptions 80116 443 10 16 CHAPTER 5 Circuit Descriptions...
Page 38: ...8 Maintenance and servicing 38 46 80116 443 10 16...
Page 40: ...9 Replacing components 40 46 80116 443 10 16...
Page 46: ...10 Spare parts 46 46 80116 443 10 16 Fig 10 6 40322 024 1 only 230 V 40322 025 1 40322 013 1...