16
CHAPTER 9
Replacing components
Safety information
Replacing components
The "wearing parts", such as seals or similar components, are designed
such that, given normal operating conditions, it should not be necessary to
replace components in the first 10 years (except for filter and consum-
ables).
The following parts may have to be replaced during servicing:
Contact pins on the
contact circuit board
Remove main filter; do not open the housing! Remove the contact pins by
pulling them out firmly by hand.
Boards, fan and other internal
components
Having removed the housing cover, they can be removed with suitable
tools. When opening and closing the housing cover, be sure to be careful
of the grounded conductor fastening on the cover.
Software
Software may only be updated by Erbe Elektromedizin using suitable pro-
gramming tools.
WARNING!
Unplug the power cord from the outlet before exchanging parts of the
unit or cleaning it.
ATTENTION!
Adjustments, technical tests, modifications, maintenance and repair
work may only be performed by Erbe or persons trained by Erbe. If the
work is not performed by trained persons, Erbe accepts no liability and
warranty rights become void.
ATTENTION!
Only use original Erbe spare parts. The manufacturer accepts no liability
and the warranty rights becomes void if original spare parts are not
used.
Summary of Contents for IES 2
Page 1: ...10 16 80116 443 SERVICE MANUAL IES 2 1 3 x...
Page 2: ......
Page 3: ...SERVICE MANUAL IES 2...
Page 18: ...4 Installation 18 46 80116 443 10 16...
Page 19: ...19 46 5 Circuit Descriptions 80116 443 10 16 CHAPTER 5 Circuit Descriptions...
Page 38: ...8 Maintenance and servicing 38 46 80116 443 10 16...
Page 40: ...9 Replacing components 40 46 80116 443 10 16...
Page 46: ...10 Spare parts 46 46 80116 443 10 16 Fig 10 6 40322 024 1 only 230 V 40322 025 1 40322 013 1...