15
/ 46
3
•
Technical data
801
16-
443
10.
16
CHAPTER 3
Technical data
Power connection
Rated supply voltage
230 V ± 10 %
120 V ± 10 %
100 V ± 10 %
Rated supply frequency
50 / 60 Hz
50 / 60 Hz
50 / 60 Hz
Current consumption
3.0 A
4.2 A
5.0 A
Power consumption
410 watts
370 watts
375 watts
Terminal for potential equalization
yes
yes
yes
Power fuses
T 3.15 A H / 250 V
T 8 A H / 250 V
T 8 A H / 250 V
Device parameters
230 V variant
100 / 120 V variant
Suction capacity
System: 1600 l/min ± 10 %
System with new filter: >550 l/min
System: 1600 l/min ± 10 %
System with new filter: >450 l/min
Filter parameters
ULPA, retention = 99.9999 % @ 0.01
μ
m DIN EN 1822-1 Filter Class H16
Dimensions and weight
Width x height x depth (all versions)
205 x 260 x 404 mm (without main filter)
Weight (230 V version)
10.7 kg (with main filter)
Weight (100 V / 120 V version)
14.1 kg (with main filter)
Ambient conditions for transport and storage of unit
Temperature
-40 °C to +70 °C
Relative humidity
10 % – 95 %
Air pressure
54 kPa - 106 kPa
Summary of Contents for IES 2
Page 1: ...10 16 80116 443 SERVICE MANUAL IES 2 1 3 x...
Page 2: ......
Page 3: ...SERVICE MANUAL IES 2...
Page 18: ...4 Installation 18 46 80116 443 10 16...
Page 19: ...19 46 5 Circuit Descriptions 80116 443 10 16 CHAPTER 5 Circuit Descriptions...
Page 38: ...8 Maintenance and servicing 38 46 80116 443 10 16...
Page 40: ...9 Replacing components 40 46 80116 443 10 16...
Page 46: ...10 Spare parts 46 46 80116 443 10 16 Fig 10 6 40322 024 1 only 230 V 40322 025 1 40322 013 1...