9
•
Maintenance and servicing
34
/ 56
801
16-
261
201
8-
07
Technical safety check
For simplification the device to be tested is referred to below as the "test
specimen".
Important information
WARNING!
For safety reasons (personnel protection) the test specimen should
generally be operated by a suitable isolating transformer.
An exceptional case is the tests for grounded conductor resistance,
ground leakage current, and patient leakage current, in which the test
specimen is supplied with current via the safety tester.
ATTENTION!
In the event of a fault occurring in the test specimen or individual com-
ponents during the technical safety check the test steps taken so far no
longer apply.
Remedy the defect and repeat the technical safety check from the
beginning.
IMPORTANT!
It is assumed that the user knows how to operate the test specimen,
the test equipment, the measuring equipment, and auxiliary test equip-
ment. The test instructions only apply in conjunction with the relevant
test steps.
IMPORTANT!
Test equipment, measuring equipment, and auxiliary test equipment
(cables, test boxes, etc.) are listed separately at the beginning of each
test unit. Where Erbe article numbers are specified, only original Erbe
test equipment, measuring equipment, and auxiliary test equipment
may be used.
IMPORTANT!
The test report for the technical safety check can be requested from
Erbe Technical Service Tübingen. For the address see address sheet on
last page.
Summary of Contents for EIP 2
Page 1: ...2018 07 80116 261 SERVICE MANUAL EIP 2 V 1 0 x V 1 1 x ELECTROSURGERY ...
Page 2: ......
Page 3: ...SERVICE MANUAL EIP 2 ...
Page 10: ...2 Modifications 10 56 80116 261 2018 07 ...
Page 18: ...5 Installation 18 56 80116 261 2018 07 ...
Page 19: ...19 56 6 Circuit Descriptions 80116 261 2018 07 CHAPTER 6 Circuit Descriptions ...
Page 24: ...6 Circuit Descriptions 24 56 80116 261 2018 07 ...
Page 46: ...10 Replacing components 46 56 80116 261 2018 07 ...