Unit description
801
16-
261
201
8-
07
CHAPTER 3
Unit description
Unit variants
The EIP 2 (Erbe Irrigation Pump) is available in the following unit variants:
•
230 V
with ECB socket
1
•
230 V
without ECB socket
•
120 V
without ECB socket
•
100 V
with ECB socket
•
100 V
without ECB socket
Factory setting
The EIP 2 has the following settings when it is delivered by the manufac-
turer:
•
Quick activation button I = 30%
2
•
Quick activation button II = 50%
•
Quick activation button III = 80%
•
Automatic pressure control activated
for units up to software version 1.0.0
•
Automatic pressure control deactivated
(for units from software version 1.1.0)
1. ECB = Erbe Communication Bus for connecting the EIP 2 to a VIO electrosurgical unit.
2. Flow in % relative to the maximum flow of 500 ml/min.
IMPORTANT!
Once settings have been changed and saved they still apply even after
the unit has been restarted. A reset (= switching the unit back to the
factory setting) is not possible.