22
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter “Validated procedures
in overview“.
1. Prepare a cleaning bath and a separate disinfectant bath.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Rinse the surfaces with sterile, deionized water (at least
1 minute).
5. Repeat the preceding cleaning steps if the product still
shows visible contamination.
6. Immerse the product completely in a disinfectant bath, en-
suring that the product does not come into contact with oth-
er components in the bath.
7. Rinse the surfaces with sterile, deionized water (at least
1 minute).
8. Dry the product until there are no more fluid residues visible
(e.g. with filtered compressed air).
7.7 Cleaning and disinfection by machine
The washer/disinfector must have fundamentally certified ef-
fectiveness (e.g. CE marking according to DIN EN ISO 15883).
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter “Validated procedures
in overview“.
OBJ_DOKU-26720-006.fm Seite 22 Freitag, 7. April 2017 12:57 12
Summary of Contents for 20190-008
Page 2: ...OBJ_DOKU 26720 006 fm Seite 2 Freitag 7 April 2017 12 57 12 ...
Page 26: ...26 OBJ_DOKU 26720 006 fm Seite 26 Freitag 7 April 2017 12 57 12 ...
Page 88: ...88 OBJ_DOKU 26720 006 fm Seite 88 Freitag 7 April 2017 12 57 12 ...
Page 207: ...OBJ_DOKU 26720 006 fm Seite 207 Freitag 7 April 2017 12 57 12 ...