23
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter “Overview of validated
procedures“.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
5.2 Processing
limitation
Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
5.3 Dismantling
1. First, separate the cable from the instrument. Do not damage
the cable insulation while doing so.
5.4 Required
aids
Processing step
Items Needed
Precleaning
Soft plastic brush/soft cloth
50 ml disposable syringe with male
Luer-Lock connection
Female-to-female luer-lock adapter
(=irrigation adapter)
OBJ_DOKU-189884-007.fm Seite 23 Freitag, 11. Juni 2021 10:59 10
Summary of Contents for 20132-046
Page 2: ...OBJ_DOKU 189884 007 fm Seite 2 Freitag 11 Juni 2021 10 59 10 ...
Page 30: ...30 OBJ_DOKU 189884 007 fm Seite 30 Freitag 11 Juni 2021 10 59 10 ...
Page 46: ...46 OBJ_DOKU 189884 007 fm Seite 46 Freitag 11 Juni 2021 10 59 10 ...
Page 144: ...144 OBJ_DOKU 189884 007 fm Seite 144 Freitag 11 Juni 2021 10 59 10 ...
Page 158: ...158 OBJ_DOKU 189884 007 fm Seite 158 Freitag 11 Juni 2021 10 59 10 ...
Page 172: ...172 OBJ_DOKU 189884 007 fm Seite 172 Freitag 11 Juni 2021 10 59 10 ...
Page 200: ...200 OBJ_DOKU 189884 007 fm Seite 200 Freitag 11 Juni 2021 10 59 10 ...
Page 230: ...230 OBJ_DOKU 189884 007 fm Seite 230 Freitag 11 Juni 2021 10 59 10 ...
Page 244: ...244 OBJ_DOKU 189884 007 fm Seite 244 Freitag 11 Juni 2021 10 59 10 ...
Page 256: ...256 OBJ_DOKU 189884 007 fm Seite 256 Freitag 11 Juni 2021 10 59 10 ...
Page 269: ...OBJ_DOKU 189884 007 fm Seite 269 Freitag 11 Juni 2021 10 59 10 ...