Equil
TM
Insulin Patch Pump System
123
IEC60601-1-2: Table 206
Recommended separation distances between portable and mobile RF
communications equipment and the device
Maximum rated
output power of
transmitter
Separation distance according to frequency of transmitter (m)
150kHz~80MHz
d=1.2
√
P
80MHz~800MHz
d=1.2
√
P
800MHz~2.5GHz
d=2.3
√
P
These devices are intended for use in an environment in which radiated RF disturbances are
controlled. The user of the device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and
the device as recommended below, according to the maximum output power of the communications
equipment
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
0.01
0.1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
Summary of Contents for MTM-I
Page 1: ...Insulin Patch Pump System MTM I Insulin Patch Pump System User Manual TM...
Page 2: ......
Page 45: ...EquilTM Insulin Patch Pump System Figure 45 1 Figure 46 2 47...
Page 62: ...EquilTM Insulin Patch Pump System 57 Figure 70...
Page 65: ...EquilTM Insulin Patch Pump System Figure 73 3 Figure 74 4 XXXXXX...
Page 67: ...EquilTM Insulin Patch Pump System Figure 77 6 Note...
Page 68: ...EquilTM Insulin Patch Pump System 63 Figure 78...
Page 71: ...EquilTM Insulin Patch Pump System...
Page 73: ...EquilTM Insulin Patch Pump System 4...
Page 85: ...EquilTM Insulin Patch Pump System Figure 100 3 Note...
Page 87: ...EquilTM Insulin Patch Pump System Figure 101 10 10 Settings Figure 102...
Page 107: ...EquilTM Insulin Patch Pump System Figure 119 Figure 120 Note...
Page 119: ...EquilTM Insulin Patch Pump System 17 11 Storage 17 12 Other Considerations...
Page 129: ...EquilTM Insulin Patch Pump System 19 Appendix 19 1 Symbols LOT SN IVD CTRL STERILE EO IPX4...