18.4 Bolus Calculator
18.5 Bolus Delivery
18.6 Infusion Precision
At 1U/h delivery rate, the error was
measured at 0.4% as shown in Figure 122.
Figure 122
Bolus Steps
Volume
per Step
Time Interval
Between Steps
Infusion Speed
per Minute
0.05 U
0.5 μL
1s
3 U
The above test results were achieved
using pump serial number 0A001 and
reservoir batch number G13B25001
Note
Attribute
Specification
On/Off, default is off
1-150 g carb/U in 1 g carb/U increments, no default
0.1-16.7 mmol/L/U in 0.1 steps, no default value
On/Off, Default is On
2-6 hrs in 0.5hr increments,no default value
Bolus Calculator
Carb Ratio
Insulin Sensitivity Factor
Negative Correction
Active Insulin Time
Summary of Contents for MTM-I
Page 1: ...Insulin Patch Pump System MTM I Insulin Patch Pump System User Manual TM...
Page 2: ......
Page 45: ...EquilTM Insulin Patch Pump System Figure 45 1 Figure 46 2 47...
Page 62: ...EquilTM Insulin Patch Pump System 57 Figure 70...
Page 65: ...EquilTM Insulin Patch Pump System Figure 73 3 Figure 74 4 XXXXXX...
Page 67: ...EquilTM Insulin Patch Pump System Figure 77 6 Note...
Page 68: ...EquilTM Insulin Patch Pump System 63 Figure 78...
Page 71: ...EquilTM Insulin Patch Pump System...
Page 73: ...EquilTM Insulin Patch Pump System 4...
Page 85: ...EquilTM Insulin Patch Pump System Figure 100 3 Note...
Page 87: ...EquilTM Insulin Patch Pump System Figure 101 10 10 Settings Figure 102...
Page 107: ...EquilTM Insulin Patch Pump System Figure 119 Figure 120 Note...
Page 119: ...EquilTM Insulin Patch Pump System 17 11 Storage 17 12 Other Considerations...
Page 129: ...EquilTM Insulin Patch Pump System 19 Appendix 19 1 Symbols LOT SN IVD CTRL STERILE EO IPX4...