Krautkramer USN 52R/USN 52L
Issue 05, 02/00
5-11
Operation:
The
MULTECHO
setting of the function
MEASURE
(lower operating level) is a prerequisite for operation.
• Simultaneously press
N
and
O
in
a-WIDTH
or
a-THRESH
in order to switch to the settings for
Gate b.
a-THRESH Determination of threshold
With this function you determine the threshold for the
selected gate. You can set a range from 10% to 90%
screen height in 1% steps.
Operation:
– Set the required threshold value in
a-THRESH
with
M
.
The selected gate is displayed as a line.
If there is a violation of the set threshold (signals ex-
ceeding or falling below it), an alarm is activated.
The exact gate position is defined by two further
parameters:
• Gate start
(a-START)
• Gate width
(a-WIDTH)
a-START Start of the gate
In the function
a-START
you determine the start of
Gate a within a range from 0 to 5000 mm (continuous).
a-WIDTH Width of the gate
In the function
a-WIDTH
you determine the width within
a range from 1 to 5000 mm (continuous).
Operation:
– Set the gate position in
a-START
and
a-WIDTH
with the assigned
M
keys.
Adjustment of the USN 52R/USN 52L
Operation
Krautkramer USN 52R/USN 52L
Issue 05, 02/00
5-11
Operation:
The
MULTECHO
setting of the function
MEASURE
(lower operating level) is a prerequisite for operation.
• Simultaneously press
N
and
O
in
a-WIDTH
or
a-THRESH
in order to switch to the settings for
Gate b.
a-THRESH Determination of threshold
With this function you determine the threshold for the
selected gate. You can set a range from 10% to 90%
screen height in 1% steps.
Operation:
– Set the required threshold value in
a-THRESH
with
M
.
The selected gate is displayed as a line.
If there is a violation of the set threshold (signals ex-
ceeding or falling below it), an alarm is activated.
The exact gate position is defined by two further
parameters:
• Gate start
(a-START)
• Gate width
(a-WIDTH)
a-START Start of the gate
In the function
a-START
you determine the start of
Gate a within a range from 0 to 5000 mm (continuous).
a-WIDTH Width of the gate
In the function
a-WIDTH
you determine the width within
a range from 1 to 5000 mm (continuous).
Operation:
– Set the gate position in
a-START
and
a-WIDTH
with the assigned
M
keys.
Adjustment of the USN 52R/USN 52L
Operation
Summary of Contents for USN 52L
Page 7: ...0 6 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 8: ...Krautkramer USN 52R USN 52L Issue 05 02 00 1 1 Introduction 1 Introduction 1 ...
Page 19: ...1 12 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 25: ...2 6 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 32: ...Krautkramer USN 52R USN 52L Issue 05 02 00 4 1 Basics of operation 4 Basics of operation 4 ...
Page 45: ...4 14 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 46: ...Krautkramer USN 52R USN 52L Issue 05 02 00 5 1 Operation 5 Operation 5 ...
Page 117: ...5 72 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 118: ...Krautkramer USN 52R USN 52L Issue 05 02 00 6 1 Documentation 6 Documentation 6 ...
Page 123: ...6 6 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 127: ...7 4 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 150: ...Krautkramer USN 52R USN 52L Issue 05 02 00 9 1 Specifications 9 Specifications 9 ...
Page 155: ...9 6 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 156: ...Krautkramer USN 52R USN 52L Issue 05 02 00 10 1 Appendix 10 Appendix 10 ...
Page 165: ...10 10 Issue 05 02 00 Krautkramer USN 52R USN 52L ...
Page 166: ...Krautkramer USN 52R USN 52L Issue 05 02 00 11 1 Changes 11 Changes 11 ...
Page 168: ...Krautkramer USN 52R Issue 05 02 00 12 1 Index 12 Index 12 ...
Page 176: ...Option DGS Operating Manual ...
Page 179: ...DGS 4 Issue 02 03 99 Krautkramer USN 52R USN 52L Option DGS ...
Page 180: ...Krautkramer USN 52R USN 52L Option DGS Issue 02 03 99 DGS 5 Introduction 1 Introduction 1 ...
Page 191: ...DGS 16 Issue 02 03 99 Krautkramer USN 52R USN 52L Option DGS ...
Page 203: ...DGS 28 Issue 02 03 99 Krautkramer USN 52R USN 52L Option DGS ...
Page 207: ...DGS 32 Issue 02 03 99 Krautkramer USN 5 USN 52L Option DGS ...