![Epoch WT-PP User Manual Download Page 5](http://html1.mh-extra.com/html/epoch/wt-pp/wt-pp_user-manual_2418062005.webp)
Contacts us : email :
, Ph: 080-23481364
LOAD CALIBRATION
Calibration accesses through hardware lock protection. Please short pin no. 7
and 8 in rear panel to enable calibration .
Accessing menus requires entering the default factory set valid 3-digit password i.e. ‘000’. It
is suggested that you change the password.
This menu will help you to calibrate load.To access load calibration menu.
S.No
Instruction
Display
1.
Press
MENU/ESC
and enter the default factory
set password ‘000’ and press
ENTER/TARE
.
2.
Navigate to
“AUto” or “nAnUL”
and press
ENTER/TARE
.
Two different Load calibrations available
“AUTO” & “MANUAL”. Difference, in manual
calibration we have to enter the span count
(Known weight count).
3.
Navigate to
Maximum Capacity (n-CAP)
and
press
ENTER/TARE
.
Enter the max capacity of load cell within the
range (0.01-999.99). E.g., ‘020.00’
4.
Navigate to
Decim
al
Point (d-Pt)
and press
ENTER/TARE
.
Set the decimal value with the possible values
of (0, 1, 2, 3).
You can enter the values by pressing “
UP
” or “
SHIFT
” keys.
Press
ENTER/TARE
key to store the current value.
Press
MENU/ESC
key to either return to previous menu or cancel the entered
values.
000
CALIb
n-CAP
020.00
d-Pt
2
AUto
NAnUL