COMMUNICATION SETUP
d-Id
01
bAUd
9600
C-nod
n-bUS
OFFSE
00000
To access Serial communication setup menu, navigate to menu ->S-Conand press
ENTER/TARE
.
S.No.
Instruction
Display
1.
Navigate to
Device ID (d-Id)
and press
ENTER/TARE
.
Enter the Device ID (1-99). It is required to identify the
system, in case of multiple systems are deployed.
2.
Navigate to
Baud Rate (bAUd)
and press
ENTER/TARE
.
It is used to set the communication between unit and
host device. It is settable between 2400, 4800, 9600,
19200, 38400, 57600, 115200bps.
3.
Navigate to
Communication mode (C-nod)
and press
ENTER/TARE
.
Available modes are,
o
Request mode
o
Dump mode
o
ModBusRTU mode
o
ModBus ASCII mode
Refer Serial Communication protocol section.
4.
Navigate to OFFSET(OFFSE) and press
ENTER/TARE.
User can enter the OFFSET value here with range of
(00000-49999).