Error Messages
S.No.
Error
Code
Description
1.
-oC-
If the value of current weight or displacement exceeds the
[Maximum load capacity or Maximum displacement value +
(9 * resolution)], then “OC” is displayed.
2.
oPEn
If the value of load cell exceeds its range or if the load cell
disconnected, “or” is displayed.
Or
If the value of load cell under ranged, “ur” is displayed.
3.
PASEr
If the entered password was wrong then this error message
will display. Please enter the correct password.
4.
U-Err
If the entered value exceeds the limit then this error
message will display. Please enter the correct value.
5.
S-Err
If the entered cal span is not more than the cal zero + 50
value then the system will display this error message.
6.
C-Err
If calibration back not taken then this error message will
display.
7.
C-dIS
Calibration menu access disabled.