PROBLEMS, CAUSES AND SOLUTIONS
The device is not turning on:
• Make sure the plug has been correctly inserted in the main supply
• Make sure switch is on the position turned on (I)
The device is turned on but it’s not nebulising:
• Make sure to have inserted the pisper into the medicine cup (8)
• Make sure the air connection pipe is not bended or crushed (9)
• Make sure air suction filter is not stopped or dirty .In that case replace it with a new one (10)
• Make sure the medicine cup has been filled in with medicine.
Note: If the device shall not take again his correct functioning in spite of inspections effected, contact the After Sales
Service.
■
MANUAL AND DECLARATIONS OF THE CONSTRUCTOR – ELECTROMAGNETICS EMISSIONS
The appliance for aerosol therapy can be use in the electromagnetic setting here indicated:
Emission test
Compliance
Manual to the electromagnetic setting
Irradiate/conducted
emission
Group1
The disposal used the energy RF only for its internal function. So its
RF emissions are very low and they do not cause any interferences to
other electrical appliances.
Irradiate/conducted
emissions CISPR11
Harmonic emissions
IEC/EN 61000-3-2
voltage/flicker
fluctuations
IEC/EN 61000-3-3
Class (B)
Class (A)
The appliance is adapt to be used in all the setting ,include the
domestic environment and all the settings connecting to the
distribution public net which give alimentation to the environment
used for domestic use
Electrostatic
discharges( ESD)
IEC/EN 61000-4-2
±6kV in contact
±8kV in air
The floors should be done in wood, cement or ceramics. If the floors
are cover by synthetic material, the relative humidity should be as a
maximum of 30%
speed/burst transitors
IEC/EN 61000-4-4
±2kV power
The supply should de typical of an hospital or a commercial setting
Surge
IEC/EN 61000-4-5
±1kV
differential
mode
The supply should de typical of an hospital or a commercial setting
Voltage hole, short
breaks and voltage
change
IEC/EN 61000-4-11
<5%Ut for 0.5
cycle
40%Ut for 05
cycle
70%Ut for 25
cycle
<5%Ut for 5
sec
The supply should de typical of an hospital or a commercial setting If
the user need a continuing working of the device, it’s recommended to
use the appliance under a continuity group.
Magnetic field
IEC/EN 61000-4-8
3A/m
The magnetic field should be typical of an hospital or an commercial
setting.
Conducts immunity
IEC/EN 61000-4-6
3Vrms 150kHz
to 80 MHz 8
(for appliances
life-supporting)
Irradiate immunity
IEC/EN 61000-4-3
3V/m 80 MHz
to 2.5GHz ( for
appliances
NOT life-
equipment)
Note
UT is the value of the supply voltage
Elettromedical devices require special care. During installation and use with respect to EMC requirements, it therefore required that
they are installed and/or used according to the manufacturer’s specification. Potential risk of electromagnetic interference with other
devices, in particular with other devices for analysis and treatment. Radio and mobile telecommunications devices or portable RF
(mobile phones or wireless connections) may interfere with the operation of elettromedical device.
Manufactured by: EP S.p.A. via Del Commercio 1 Travagliato (BS) ITALY