![Environmental Express Oakton 300 Series Instruction Manual Download Page 48](http://html.mh-extra.com/html/environmental-express/oakton-300-series/oakton-300-series_instruction-manual_2416084048.webp)
COND, TDS, Sal Setup
37
Default settings are in
Bold
.
Parameter
Settings
Details
TDS Setup
Use TDS Setup to customize settings for the EC
sensor head.
Factor Type
Linear
Linear factor with adjustable factor from 0.40 to
1.00
442
Myron L 442: refer to Natural water (40%
Sodium sulfate, 40% Sodium carbonate, 20%
Sodium chloride)
EN27888
European environmental standard
NaCl
NaCl Salinity curve
Factor Value
0.40 - 10.0
Default:
0.50
Enter the factor value used when the Factor Type
Linear setting is active.
Unit
mg/L
ppm
Set the unit of the TDS measurement value.
Sal Setup
Use Sal Setup to customize settings for the EC
sensor head.
Unit
PPT
%
Set the unit of the salinity measurement value.
Sal Type
Set the type of salinity measurement.
NaCl
NaCl: Compensation by NaCl concentration
Seawater
Seawater: Compensation by Practical Salinity
Scale (UNESCO 1978)
CAL Data Clear
Erase the calibration data.
Cell Type
4-Cell
2-Cell
Set the type of EC sensor cartridge as 4-Cell or 2-
Cell type.
Cartridge Info.
Use the Cartridge Information to update meter
settings for the model and SN of the pH sensor
cartridge.
Model
Enter a model name using the alphanumeric entry
screen (up to 10 characters)
SN
Enter a serial number using the alphanumeric
entry screen (up to 10 characters)
General Setup
refer to General
Setup on page
33
Use General Setup to customize settings for the
meter.
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30
Summary of Contents for Oakton 300 Series
Page 1: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Page 4: ...Regulations III www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Page 11: ...Contents X www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...
Page 62: ...www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...