DO Calibration
21
Calibration
1. Press the
key. Meter starts checking various calibration values with a blinking
on screen.
*Switch to High [100 %] and Low [0 %] calibration mode by pressing the
key in
DO calibration mode.
2. Wait for the to stabilize (stable calibration reading).
3. Press the key to confirm calibration data.
4. Meter displays
DONE
indicating end of the DO calibration procedure.
5. Switch to the measurement screen. For the second calibration, repeat the same cali-
bration procedure from 1.