EC Calibration
17
Calibration
1. After placing the EC sensor in the standard solution, press the
key.
2. Meter starts checking various calibration values with a blinking on screen.
* In the buffer setting for CUSTOM calibration, press
key and then adjust the
calibration value.
3. Wait for the to stabilize (stable calibration reading).
4. Press the key to confirm calibration data.
5. Meter displays
DONE
indicating end of the conductivity calibration procedure.
6. Switch to the measurement screen. For the second calibration, repeat the same cali-
bration procedure from 1.