12
9 Stimulator Output Parameters
Wave form
Biphasic asymmetrical TENS
Pulse Width
50, 100, 150, 200 and 250 µs
Pulse frequency
1, 2, 3, 4, 5, 10, 12, 14, 16, 18, 20, 25, 30, 35, 40, 45, 50, 60, 70, 80, 90,
100, 110, 120, 130, 140 and 150 Hz
Frequency modulation
Program I, 20Hz to 100 Hz and back in 12 seconds
Program M, 1Hz to 150Hz
Intensity modulation
Program G, intensity down to 70% and back over 10 seconds
Pulse Width modulation
Program H, pulse width modulation from 250µs to 50µs and back in 12
seconds
Burst
2 Hz (0,25 seconds at 100Hz)
Amplitude
0 – 90 mA
Treatment time
10 – 20 – 30 – 45 – 60 minutes and Constant
10 Illustrations
10.1 Asymmetrical biphasic TENS waveform
t
Pulse width
f
p
Pulse frequency
10.2 Asymmetrical
biphasic
burst
TENS
waveform
f
B
Burst frequency
Summary of Contents for TensMed 911
Page 1: ...TensMed 911 User Manual...
Page 3: ...1 TensMed 911 User Manual...