Enraf Nonius TensMed 911 User Manual Download Page 5

3

1 Product Description

The TensMed 911 is a 2-channel electrotherapy apparatus, that has optimized pain management, using
asymmetric TENS current waveforms.

The apparatus provides both protocol driven and manual operation.

Protocol driven operation provides both factory or user defined treatments. In this mode the 2 channels
are linked, i.e. parameter settings apply to all channels simultaneously, the treatment timer included.

With manual operation the channels are linked, the current amplitude can be set individually on each
channel.

WARNING (USA only): The TensMed 911 is a prescription device that should only be used under the
supervision or by the order of a physician or other licensed healthcare provider.

2 Foreword

2.1

This manual

This manual has been written for the owners and operators of the TensMed 911. It contains general
instructions on operation, precautionary practices, maintenance and parts information. In order to
maximize use, efficiency and the lifetime of your unit, please read this manual thoroughly and become
familiar with the controls as well as the accessories before operating the unit.
Specifications put forth in this manual were in effect at the time of publication. However, owing to Enraf-
Nonius BV policy of continual improvement, changes to these specifications may be made at any time
without obligation on the part of Enraf-Nonius BV.

2.2

Product liability

A law on Product Liability has become effective in many countries. This Product Liability law implies,
amongst other things, that once a period of 10 years has elapsed after a product has been brought into
circulation, the manufacturer can no longer be held responsible for possible shortcomings of the product.

2.3

Limitations of liability

To the maximum extent permitted by applicable law, in no event will Enraf-Nonius or its suppliers or
resellers be liable for any indirect, special, incidental or consequential damages arising from the use of or
inability to use the product, including, without limitation, damages for loss of goodwill, work and
productivity, computer failure or malfunction, or any and all other commercial damages or losses, even if
advised of the possibility thereof, and regardless of the legal or equitable theory (contract, tort or
otherwise) upon which the claim is based. In any case, Enraf-Nonius’s entire liability under any provision
of this agreement shall not exceed in the aggregate the sum of the fees paid for this product and fees for
support of the product received by Enraf-Nonius under a separate support agreement (if any), with the
exception of death or personal injury caused by the negligence of Enraf-Nonius to the extent applicable
law prohibits the limitation of damages in such cases.
Enraf-Nonius can not be held liable for any consequence resulting from incorrect information provided by
its personnel, or errors incorporated in this manual and / or other accompanying documentation (including
commercial documentation)
The opposing party (product’s user or its representative) shall disclaim Enraf-Nonius from all claims
arising from third parties, whatever nature or whatever relationship to the opposing party.

Summary of Contents for TensMed 911

Page 1: ...TensMed 911 User Manual...

Page 2: ...Copyright Enraf Nonius B V P O Box 12080 3004 GB Rotterdam The Netherlands Tel 31 0 10 20 30 600 Fax 31 0 10 20 30 699 info enraf nonius com www enraf nonius com Part number 1427 755 41 August 2005...

Page 3: ...1 TensMed 911 User Manual...

Page 4: ...c effects 7 7 4 Current density 7 7 5 Connection and disconnection reactions 7 8 Operating Instructions 8 8 1 Operator Controls 8 8 2 Intensity adjustment 9 8 3 Safety features 9 8 4 Additional featur...

Page 5: ...ct Liability law implies amongst other things that once a period of 10 years has elapsed after a product has been brought into circulation the manufacturer can no longer be held responsible for possib...

Page 6: ...with air oxygen or nitrous oxide This device should be kept out of the reach of children CAUTION Read understand and practice the precautionary and operating instructions Know the limitations and haza...

Page 7: ...ing pregnancy Stimulation should not be applied over swollen infected or inflamed areas of skin eruptions e g phlebitis thrombophlebitis varicose veins etc See also chapter 3 Precautionary Instruction...

Page 8: ...0mm 1 pack 10 sheets of 4 pc 3444057 EN Trodes self adhesive electrodes size 50x50mm 1 pack 10 sheets of 4 pc 3444058 EN Trodes self adhesive electrodes size 50x90mm 1 pack 10 sheets of 4 pc 3444146 E...

Page 9: ...sponges to keep the effects to a minimum Make sure that the sponges are kept well moistened and place the thick side of the sponge between the flexible rubber electrodes and the patient 7 4 Current de...

Page 10: ...ters that can be selected are Waveform Frequency Pulse Width Constant or Burst Timer The PAUSE function interrupts a Program which is using the Timer It only operates when a program is running with In...

Page 11: ...is a low battery detector in the unit The LCD screen will flash the symbol of a battery when the batteries need replacing 8 5 Usage monitor USAGE MONITOR For Professional Use A Usage Monitor is includ...

Page 12: ...E Pulsing Tapping Chronic pain nausea travel and morning sickness Avoid use on traumatised muscles F Pulsing Tapping Most types of acute and chronic pain G Varies Acute and chronic pain H Varies Acut...

Page 13: ...locking 8 8 User Maintenance 8 8 1 Cleaning of apparatus To clean the unit turn it off Clean the unit with a damp cloth Do not use abrasive cleaners A small amount of mild household detergent may be u...

Page 14: ...n 12 seconds Program M 1Hz to 150Hz Intensity modulation Program G intensity down to 70 and back over 10 seconds Pulse Width modulation Program H pulse width modulation from 250 s to 50 s and back in...

Page 15: ...lass according to IEC 60601 1 type BF Dimensions 144 x 80 x 27 5 mm h x w x d Weight 180 gram included batteries Environmental conditions for transport and storage Temperature 10 to 50 C Relative humi...