background image

  

 

EX-5155 

ENMET 

Addendum: Ammonia Calibration 

Calibration Procedure for EX-5155-MOSAmmonia 

Manual Release Date –

 November 26,  

2018

 

Manual Part No. –

 

 80003-098 

P a g e  

 20 

All MOS sensors require at least 15% RH for proper operation.  RH of 50% or greater is ideal for MOS sensor calibration.  The  

common technique for humidifying dry gas from a cylinder is to bubble it through a humidifier assembly.  However, ammonia is  

readily absorbed in water and therefore the standard regulator assembly cannot be used.  

ENMET

 has developed a special calibration adapter for ammonia that allows the calibration gas to be humidified without passing  

through water.  The part number of this adapter is 03700-005.  Instead of the calibration gas being flown through the water the 

calibration gas is passed over the water.  The calibration gas picks up enough humidity to properly calibrate the sensor, but not enough 

to absorb all the ammonia calibration gas 
For 0 – 500 ppm range ammonia transmitters, the proper calibration point is 200 ppm.  The calibration gas cylinder is 300 ppm NH3 in 

air.  While flowing the 300 ppm ammonia through the calibration adapter humidifier assembly 100 ppm of the ammonia is absorbed in 

the water with the remaining 200 ppm reaching the sensor. 

 

Unscrew the bowl from the humidifier and place enough water into the bowl to reach the bottom edge of the tube. 

DO NOT let the water cover the tube.  The tube should not penetrate the water.  Connect the calibration adaptor to the cylinder 

of ammonia.  Place the unit into calibration mode (reference Section 5.0) and perform a ZERO calibration.  DO NOT enter the 

span sequence. 

 

 

Open the valve on the calibration adapter regulator.   

 

 

Verify that the gas sample is passing over the water by looking for gentle ripples on the water surface.  There should 

not be bubbling.  Flow rate is preset by the regulator and needs no adjustment.   

 

 

Connect the calibration cup to the sensor and allow the gas to flow for 3 minutes.   

 

 

At the end of the 3 minutes tap the SELECT switch on the 

EX-5155-MOS

 to start the span operation.  The display will 

alternate between PV and a number.   

 

 

Continue to let the calibration gas flow for an additional 1 minute then tap the SELECT switch again.  This will end the 

span operation and the display should read CAL OK.   

 

 

Tap the MENU switch until EXIT appears and then tap the SELECT switch to exit the calibration menu and return the 

unit to its operation mode. 

The 

EX-5155-MOS 

Ammonia sensor requires special calibration techniques that are not common with other gases. 

 

 

 

Notes: 

Summary of Contents for EX-5155-MOS

Page 1: ...ENMET 680 Fairfield Court Ann Arbor MI 48108 734 761 1270 Fax 734 761 3220 www enmet com EX 5155 MOS Toxic Gas Sensor Transmitter Operation Manual ...

Page 2: ...EX 5155 ENMET Manual Release Date November 26 2018 Manual Part No 80003 098 P a g e 1 ...

Page 3: ...r Replacement 16 6 0 REPLACEMENT PART NUMBERS 17 List of Figures and Tables Figure 1 EX 5155 MOS Features 5 Figure 2 EX 5155 MOS Mounting Dimensions 6 Figure 3 Terminal Positions EX 5155 MOS Sensor Transmitter 8 Table 1 EX 5155 MOS maintenance Menus Sequence 11 Figure 4 Calibration Adapter EX 5155 MOS Sensor Transmitter 12 Table 3 Examples of Standard and Non Standard Calibration Gas 13 Table 2 Ca...

Page 4: ...EX 5155 ENMET Manual Release Date November 26 2018 Manual Part No 80003 098 P a g e 3 ...

Page 5: ...ction and after you have received written acknowledgment of the damage from the carrier contact ENMET Customer Service for return authorization and further instructions Have your Purchase Order and Sales Order numbers available ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage usually 100 00 Repair or replacement charges above that ...

Page 6: ...ic switches control the instrument maintenance functions The switch locations are indicated by MENU and SELECT A magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and then removing it Referred to as tap Since the magnetic field penetrates the window the enclosure cover is not removed in order to perform calibration Three alarm points are prepro...

Page 7: ...ansmitters Also take into consideration environmental factors when deciding on S T location Avoid locations where the S T may be damaged by liquid immersion excessive heat or other know hazards Also take precautions to insure condensation insid of the conduit does not enter the S T 3 1 Mounting the EX 5155 MOS Enclosure Mount the enclosure using the two mounting holes provided see Figure 2 Pay par...

Page 8: ...Do not damage the magnetic switches Remove the two overlay standoffs and remove the circuit board exposing the terminal strips on the bottom of the circuit boar d Do not disconnect the circuit board wiring Connect the wires from the controller power supply to the supplied J4 plug then attach to J4 terminal Connect the wires from the sensor to the supplied J8 plug then attach to the J8 terminal See...

Page 9: ...closure Cutaway View Display Overlay Magnetic Switches 2 places Printed Circuit Board PCB Figure 3 Terminal Positions EX 5155 MOS Sensor Transmitter When wiring is complete reassemble the EX 5155 MOS Use caution when installing the overlay so as not to damage the magnetic switches Put the cover back on the S T Do Notapply power to the S T without the cover in place Manual Release Date November 26 ...

Page 10: ...ng the instrument status See the table below Purging occurs automatically for instruments with sensors that require purging Purge wil l occur for 5 minutes on instrument start up followed by a 5 minute recovery period At the end of the recovery period the transmitter should be ready for operation If additional purge time is required refer to Section 4 1 3 for more information Example of Display ma...

Page 11: ...t point Tap the magnet over the MENU button SELECT SELECT A1 A2 10 20 MENU MENU Display indicates Alarm 2 Set point Tap the magnet over the MENU button No Function for the SELECT button in this mode Display indicates Alarm 3 Set point Tap the magnet over the MENU button A3 50 SELECT SELECT MENU Display indicates mA Span range Full Scale mA 100 Tap the magnet over the MENU button MENU Display retur...

Page 12: ...agnet over MENU switch for 2 5 seconds to enter the Maintenance Menu The Power Fault LED will flash Green Red to indicate the EX 5155 MOS is in Maintenance Mode To exit the maintenance Menu and return to the Normal Exit Display Mode If intended function Tap the magnet over SELECT switch Tap the magnet over the MENU switch to advance to the Zero procedure For adjusting Zero Zero If intended functio...

Page 13: ...mitters Calibration equipment is available from ENMET Corporation to calibrate the EX 5155 MOS sensor transmitters A calibration adapter will have a fitting for the gas cylinder on one side and a cover to go over the sensor housing on the other Generally a cylinder of 20 9 Oxygen is used to provide a fresh air reference or Zero point for the calibration Another cy linder is used to provide the Spa...

Page 14: ...ibrations use for examples only Consult ENMET for specifications and availability Ammonia 0 300 PPM 150 PPM 300 PPM 300 PPM 03019 000 300 PPM Ammonia See Addendum Ammonia Calibration 1200 PPM Freon 134 Freon 134A Freon 12 0 2000 PPM 0 2000 PPM 500 PPM 500 PPM 1000 PPM 1000 PPM 1500 PPM 1500 PPM 03032 000 03032 000 2 LEL Methane 300 PPM Carbon 1200 PPM Freon 12 Monoxide 2 LEL Methane 500 PPM Hydrog...

Page 15: ...an change the Calibration Gas Level HOLD the magnet over the SELECT switch for 2 4 seconds The MENU switch changes digit indicated by underscore cursor 3 Attach the associated calibration gas cylinder to the regulator and calibration cover See to Figure 3 4 Open the valve to apply the calibration gas to the sensor 5 Watch for the signal level to stabilize Refer to Table 4for typical response times...

Page 16: ...2 2 SELECT Cal OK Apply Calibration Gas until signal value becomes stable about 1 to 4 minutes See Figure 4 OR Same mV If cal is good display will indicate OK or Same OR If cal is not within preset range display will indicate Bad Sens Bad Sens NOTE Some software revisions require the SELECT switch be tapped to accept the signal To change Alarm set points Alarm1 Λ 10 SetTDsec 0 SELECT Tap Menu swit...

Page 17: ...ng your Distributor or ENMET Corporation Technical Support To replace a sensor it is not necessary to open the transmitter housing Remove the set screw from sensor housing base Unscrew the sensor housing cover and remove spacer Note the orientation of spacer Unplug the sensor form PC Board Plug new sensor into PC Board and replace spacer Replace spacer with grooved edge toward sensor housing cover...

Page 18: ...00 03016 000 03018 000 03019 000 ConsultENMET Distributor orENMET Corp for additional sensors Regulator Overlay Magnet Calibration Sampling Adapter 03700 008 06000 058 50030 001 03700 034 Calibration Gas Consult ENMET Distributor orENMET Corp Regulator Ammonia 03700 005 Manual Release Date November 26 2018 Manual Part No 80003 098 P a g e 17 ...

Page 19: ... by ENMET for professional collection services or legal fees to collect on a customer invoice will be added to any future business conducted between ENMET and that customer 7 4 Warranty Information and Guidelines Equipment must be returned prepaid to the point of origin and ENMET will prepay the return transportation charges Transportation prepaid by ENMET will be by most economical means e g FedE...

Page 20: ... percentage of total cost expended prior to cancellation 7 6 Returning an Instrument for Service Instructions Contact the ENMET Service Department for all service requests Phone 734 761 1270 Email repair enmet com Fill out the Service Request Form found at the end of this manual and return with your instrument for all needs Please send your instrument for service to the site in which the product w...

Page 21: ...he ammonia is absorbed in the water with the remaining 200 ppm reaching the sensor Unscrew the bowl from the humidifier and place enough water into the bowl to reach the bottom edge of the tube DO NOT let the water cover the tube The tube should not penetrate the water Connect the calibration adaptor to the cylinder of ammonia Place the unit into calibration mode reference Section 5 0 and perform ...

Page 22: ......

Page 23: ...dress Contact Name Email Phone Fax PO Reference PAYMENT METHOD COD VISA MasterCard American Express Security Code Card Number Exp Date Name as it Appears on Card RETURN SHIPPING METHOD UPS Ground UPS 3 Day UPS Next Day UPS ND Air UPS 2 Day Air Select Air Saver UPS Account FedEx Ground FedEx Air FedEx Air FedEx Air 2 FedEx Air Express Saver Overnight Std Day Overnight P 1 FedEx Account Insure Shipm...

Reviews: