background image

  

EX-5155 

ENMET  

5.0 Maintenance of the EX-5155-MOS 

C

AUTION

:

 Do not open the 

EX-5155-MOS 

S/T in a classified area.  

Do Not Attempt A Span Procedure Without Calibration Gas 

 

Applied to The Sensor

; if this is done, the S/T is forced into a calibration fault mode.  

Magnetic switches control the 

M

ENU

 and 

S

ELECT 

functions.  The 

M

ENU

 and 

S

ELECT

 switch locations are indicated on the display  

panel, see 

Figure 3.

  The 

M

ENU

 switch is used to display the various menu options and make incremental changes to numbers such as  

alarm points, calibrations gas, etc.  The 

S

ELECT

 switch is used to select that option, set zero or span digit.  Most maintenance  

functions are controlled by simple taps of the supplied magnet on the transmitter glass, below the 

M

ENU 

and 

S

ELECT 

 boxes on the  

front panel.  

5.1 Maintenance Menu 

To enter the maintenance menu hold the magnet over the 

M

ENU

 switch for 2 to 4 seconds  

Table 1

 indicates the maintenance menu sequence see 

Figure 5

 for a detailed maintenance menu flow chart.  

Table 1: EX-5155-MOS maintenance Menus Sequence

 

Example of Display  

Function  

Normal Display Mode  

Measurement of target gas  

5ppm 

Hold the magnet over 

M

ENU

 switch for 2 – 5 seconds to enter the Maintenance Menu 

 

The Power/Fault LED will flash Green 

– Red to indicate the 

EX-5155-MOS is in Maintenance Mode 

 

To exit the maintenance Menu and return to the Normal 

 

Exit 

Display Mode:  

If intended function Tap the magnet over 

S

ELECT

 switch  

Tap the magnet over the 

M

ENU

 switch to advance to the Zero procedure 

 

For adjusting Zero:  

Zero 

If intended function Tap the magnet over 

S

ELECT

 switch  

Tap the magnet over the 

M

ENU

 switch to advance to the Span procedure 

 

For adjusting the Span: 

 

Span 

If intended function Tap the magnet over 

S

ELECT

 switch  

Tap the magnet ove

r the M

ENU

 switch to advance to each Alarm set point procedures 

 

For adjusting the Alarm 1, 2 and 3 set points: 

 

If Intended function Tap the magnet over 

S

ELECT

 switch  

Alarm1 

Alarm2 

Alarm3 

Tap the magnet over the 

M

ENU

 switch to advance the mA Span set point procedure 

 

For adjusting the mA Span set point: 

 

If intended function Tap the magnet over 

S

ELECT

 switch  

mA Span 

Taping the M

ENU

 switch without taping the 

S

ELECT

 switch will allow you to cycle through the menu options. 

 

You must Tap the 

S

ELECT

 switch in order to change the desired operation. 

 

N

OTE

: If the S/T fails to respond, the magnet may have become weak and may need to be replaced. 

 

Manual Release Date – 

November 26, 2018

 

Manual Part No. – 80003-098 

P a g  e 

  | 11  

Summary of Contents for EX-5155-MOS

Page 1: ...ENMET 680 Fairfield Court Ann Arbor MI 48108 734 761 1270 Fax 734 761 3220 www enmet com EX 5155 MOS Toxic Gas Sensor Transmitter Operation Manual ...

Page 2: ...EX 5155 ENMET Manual Release Date November 26 2018 Manual Part No 80003 098 P a g e 1 ...

Page 3: ...r Replacement 16 6 0 REPLACEMENT PART NUMBERS 17 List of Figures and Tables Figure 1 EX 5155 MOS Features 5 Figure 2 EX 5155 MOS Mounting Dimensions 6 Figure 3 Terminal Positions EX 5155 MOS Sensor Transmitter 8 Table 1 EX 5155 MOS maintenance Menus Sequence 11 Figure 4 Calibration Adapter EX 5155 MOS Sensor Transmitter 12 Table 3 Examples of Standard and Non Standard Calibration Gas 13 Table 2 Ca...

Page 4: ...EX 5155 ENMET Manual Release Date November 26 2018 Manual Part No 80003 098 P a g e 3 ...

Page 5: ...ction and after you have received written acknowledgment of the damage from the carrier contact ENMET Customer Service for return authorization and further instructions Have your Purchase Order and Sales Order numbers available ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage usually 100 00 Repair or replacement charges above that ...

Page 6: ...ic switches control the instrument maintenance functions The switch locations are indicated by MENU and SELECT A magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and then removing it Referred to as tap Since the magnetic field penetrates the window the enclosure cover is not removed in order to perform calibration Three alarm points are prepro...

Page 7: ...ansmitters Also take into consideration environmental factors when deciding on S T location Avoid locations where the S T may be damaged by liquid immersion excessive heat or other know hazards Also take precautions to insure condensation insid of the conduit does not enter the S T 3 1 Mounting the EX 5155 MOS Enclosure Mount the enclosure using the two mounting holes provided see Figure 2 Pay par...

Page 8: ...Do not damage the magnetic switches Remove the two overlay standoffs and remove the circuit board exposing the terminal strips on the bottom of the circuit boar d Do not disconnect the circuit board wiring Connect the wires from the controller power supply to the supplied J4 plug then attach to J4 terminal Connect the wires from the sensor to the supplied J8 plug then attach to the J8 terminal See...

Page 9: ...closure Cutaway View Display Overlay Magnetic Switches 2 places Printed Circuit Board PCB Figure 3 Terminal Positions EX 5155 MOS Sensor Transmitter When wiring is complete reassemble the EX 5155 MOS Use caution when installing the overlay so as not to damage the magnetic switches Put the cover back on the S T Do Notapply power to the S T without the cover in place Manual Release Date November 26 ...

Page 10: ...ng the instrument status See the table below Purging occurs automatically for instruments with sensors that require purging Purge wil l occur for 5 minutes on instrument start up followed by a 5 minute recovery period At the end of the recovery period the transmitter should be ready for operation If additional purge time is required refer to Section 4 1 3 for more information Example of Display ma...

Page 11: ...t point Tap the magnet over the MENU button SELECT SELECT A1 A2 10 20 MENU MENU Display indicates Alarm 2 Set point Tap the magnet over the MENU button No Function for the SELECT button in this mode Display indicates Alarm 3 Set point Tap the magnet over the MENU button A3 50 SELECT SELECT MENU Display indicates mA Span range Full Scale mA 100 Tap the magnet over the MENU button MENU Display retur...

Page 12: ...agnet over MENU switch for 2 5 seconds to enter the Maintenance Menu The Power Fault LED will flash Green Red to indicate the EX 5155 MOS is in Maintenance Mode To exit the maintenance Menu and return to the Normal Exit Display Mode If intended function Tap the magnet over SELECT switch Tap the magnet over the MENU switch to advance to the Zero procedure For adjusting Zero Zero If intended functio...

Page 13: ...mitters Calibration equipment is available from ENMET Corporation to calibrate the EX 5155 MOS sensor transmitters A calibration adapter will have a fitting for the gas cylinder on one side and a cover to go over the sensor housing on the other Generally a cylinder of 20 9 Oxygen is used to provide a fresh air reference or Zero point for the calibration Another cy linder is used to provide the Spa...

Page 14: ...ibrations use for examples only Consult ENMET for specifications and availability Ammonia 0 300 PPM 150 PPM 300 PPM 300 PPM 03019 000 300 PPM Ammonia See Addendum Ammonia Calibration 1200 PPM Freon 134 Freon 134A Freon 12 0 2000 PPM 0 2000 PPM 500 PPM 500 PPM 1000 PPM 1000 PPM 1500 PPM 1500 PPM 03032 000 03032 000 2 LEL Methane 300 PPM Carbon 1200 PPM Freon 12 Monoxide 2 LEL Methane 500 PPM Hydrog...

Page 15: ...an change the Calibration Gas Level HOLD the magnet over the SELECT switch for 2 4 seconds The MENU switch changes digit indicated by underscore cursor 3 Attach the associated calibration gas cylinder to the regulator and calibration cover See to Figure 3 4 Open the valve to apply the calibration gas to the sensor 5 Watch for the signal level to stabilize Refer to Table 4for typical response times...

Page 16: ...2 2 SELECT Cal OK Apply Calibration Gas until signal value becomes stable about 1 to 4 minutes See Figure 4 OR Same mV If cal is good display will indicate OK or Same OR If cal is not within preset range display will indicate Bad Sens Bad Sens NOTE Some software revisions require the SELECT switch be tapped to accept the signal To change Alarm set points Alarm1 Λ 10 SetTDsec 0 SELECT Tap Menu swit...

Page 17: ...ng your Distributor or ENMET Corporation Technical Support To replace a sensor it is not necessary to open the transmitter housing Remove the set screw from sensor housing base Unscrew the sensor housing cover and remove spacer Note the orientation of spacer Unplug the sensor form PC Board Plug new sensor into PC Board and replace spacer Replace spacer with grooved edge toward sensor housing cover...

Page 18: ...00 03016 000 03018 000 03019 000 ConsultENMET Distributor orENMET Corp for additional sensors Regulator Overlay Magnet Calibration Sampling Adapter 03700 008 06000 058 50030 001 03700 034 Calibration Gas Consult ENMET Distributor orENMET Corp Regulator Ammonia 03700 005 Manual Release Date November 26 2018 Manual Part No 80003 098 P a g e 17 ...

Page 19: ... by ENMET for professional collection services or legal fees to collect on a customer invoice will be added to any future business conducted between ENMET and that customer 7 4 Warranty Information and Guidelines Equipment must be returned prepaid to the point of origin and ENMET will prepay the return transportation charges Transportation prepaid by ENMET will be by most economical means e g FedE...

Page 20: ... percentage of total cost expended prior to cancellation 7 6 Returning an Instrument for Service Instructions Contact the ENMET Service Department for all service requests Phone 734 761 1270 Email repair enmet com Fill out the Service Request Form found at the end of this manual and return with your instrument for all needs Please send your instrument for service to the site in which the product w...

Page 21: ...he ammonia is absorbed in the water with the remaining 200 ppm reaching the sensor Unscrew the bowl from the humidifier and place enough water into the bowl to reach the bottom edge of the tube DO NOT let the water cover the tube The tube should not penetrate the water Connect the calibration adaptor to the cylinder of ammonia Place the unit into calibration mode reference Section 5 0 and perform ...

Page 22: ......

Page 23: ...dress Contact Name Email Phone Fax PO Reference PAYMENT METHOD COD VISA MasterCard American Express Security Code Card Number Exp Date Name as it Appears on Card RETURN SHIPPING METHOD UPS Ground UPS 3 Day UPS Next Day UPS ND Air UPS 2 Day Air Select Air Saver UPS Account FedEx Ground FedEx Air FedEx Air FedEx Air 2 FedEx Air Express Saver Overnight Std Day Overnight P 1 FedEx Account Insure Shipm...

Reviews: