5.2.3 Set Alarm “Delay On”
The alarms may be set to delay activation by 1 second increments, from 1 to 30 seconds.
W
ARNING
:
Extended alarm delays may cause operational failures, injury or death.
Factory set default value is 5 seconds.
After entering a valid password:
Press the
OPTION
switch until;
“Maintenance Menu Set Alarm Delay”
appears on display.
Press the
SELECT
switch, "
ALARM Delay Select: O2
" is displayed.
Press the
SELECT
switch; "
ALARM Delay =
T
005
" is displayed, with the underscore cursor under the left position.
Press the
SELECT
switch to move the cursor and the
OPTION
switch to lock in the correct digit and move the cursor one
digit to the right. Press the
SELECT
switch when the last digit is accepted display returns to the "
Set Alarm Delay
"
position.
Press the
OPTION
switch to continue on to the Set New Password section
Example: Set Alarm Delay Flow Chart
5.2.4 Relay Configuration
To change a relay configuration you must enter the maintenance menu. Press the
OPTION
switch until “Enter Maint Menu” is
displayed then press
SELECT
switch for the Enter Password menu. Enter the valid password as described below.
In the "
Enter Maint Menu
" position
Press the
SELECT
switch "
Enter Password
T
0
" is displayed. Press
SELECT
switch once, to move cursor to next digit,
this will be the first digit of the password.
In the
T
000
position, the underline cursor is under the left digit.
Press the
OPTION
switch to change the left digit; select the correct digit.
Press the
SELECT
switch, which locks the digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the
display is transferred to the "
Calibration
" portion of the menu. If an invalid password is inserted you are returned to the
Enter Maint Menu display.
After entering a valid password:
Press the
OPTION
switch until “Configure Alarms” is displayed
L
= Low Alarm = Alarm 1,
H
= High Alarm = Alarm 2,
B
= Both Alarms,
= No Relay linked to channel
Example: Set Relay Configuration Flow Chart
The Table 2 below shows the default relay links for 3 channel system.
Table 2: Default Relay Links
Channel 1
Channel 2
Channel 3
Channel 4
Relay 1
Low Alarm
Channel 4 Not
Available in CP-60
Default
programming of
circuit only
Relay 2
Low Alarm
Relay 3
Low Alarm
Relay 4
Spare
Spare
Spare
Relay 5
High Alarm
High Alarm
High Alarm
Relays can be linked to specific alarms.
N
OTE
:
Each operating channel must be linked to at least 1 relay.
O
S
MAINTENANCE MENU
Set Alarm Delays
S
O
S
Alarm Delay -
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O
Press Option until the channel to be Set is displayed
Select:
XX
Alarm Delay -
0000
O
= Press Option
S =
Press Select
Note:
When in maintenance mode the “_” indicates the location of the cursor, the “
I
” block indicates a non-used location.
O
S
MAINTENANCE MENU
Configure Alarms
S
O
S
Ch 1,2,3
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O
Press Option changes letter indicated by underscore cursor
R1 L
Ch 1 2 3
R2 L
!