EQ-400 Operation Manual Rev.
Error! Reference source not found.
13
Power/Control Unit (PCU)
The Power/Control Unit contains:
Laser module
Laser power supply
Control electronics
Status indicator LEDs
Operate switch
Keyswitch
Interface connectors
Cooling water connectors
External features (refer to Figure 3):
Status Indicator LEDs
These LEDs indicate the system status. The function of these indicators is shown below in
Table
2
.
LED Label
Meaning (when lit)
POWER OK
AC power is connected to the EQ-400 PCU
LAMP ON
UV Light is on
LASER ON
Laser power is ON and laser light is being produced within the Lamp
House
REMOTE
System is in remote control mode
LASER OVERTEMP
Laser internal temperature too high
LASER FAULT
Laser module internal fault
PCU FAULT
One of the following has occurred in the Power/Control Unit:
1. Controller internal temperature too high
2. Laser power not reaching setpoint
3. Internal power supply voltage low
4. Lamp failed to ignite
LAMPHOUSE OVERTEMP
Lamphouse internal temperature too high
LAMPHOUSE INTERLOCK
One of the following has occurred in the Lamp House module:
1. Control Cable not connected properly
2. Laser fiber not connected properly
EXTERNAL INTERLOCK
External interlock open
COVER INTERLOCK
PCU cover(s) not in place
Table 2:
Status Indicator LED Functions